Manual

66 AEJ_N/AES_N-BA-e-1321
13.6 Pipette calibration
13.6.1 Entering parameters
Prior to measuring you will have to enter the parameters below.
Parameters Description
Set n Number of measurements per test volume, selectable from 5 -15
measurements
Set t
Test temperature, selectable from 0.4 25.5°C
Set P
Air pressure test chamber, selectable from 800 1050 hPa
Set h
Air humidity test chamber, selectable from 10 95 %
For devices with a fixed volume, test volume equals nominal volume.
For devices with variable volume, the following three volumes need to
be tested:
V
0
The lower limit for the effective volume or 10 % of the nominal volume
[µl], depending on whichever is the larger.
V
1/2
About 50 % of the nominal volume [µl]
V
1
Nominal volume [µl]