Operator Manual Firmware Ver.3.05 and higher Instruction Manual

Table Of Contents
BOP-1K-GL 022814 4-11/(4-12 Blank)
The Master calibration should never be overwritten. Factory, Master, and First are set to the
same values when the BOP is calibrated at the factory. The Working calibration is the active cal-
ibration. Each time a CAL:SAV is executed, the values are saved in the Working (active) area.
At the same time, the values previously stored in Working are moved to Prior, and the values
previously stored in Prior are moved to Oldest. The values previously stored in Oldest are no
longer available. Table 4-5 illustrates calibration storage and the use of the CAL:DUMP? query
and CAL:COPY command.
TABLE 4-5. CALIBRATION STORAGE
COMMAND WORKing PRIor OLDest FACTory MASTer FIRst
1. CAL:SAVE Cal 1 values Factory cal. Master cal. First cal.
2. CAL:SAVE Cal 2 values Cal 1 values No Change No Change No Change
3. CAL:SAVE Cal 3 values Cal 2 values Cal 1 values No Change No Change No Change
4. CAL:SAVE Cal 4 values Cal 3 values Cal 2 values No Change No Change No Change
5. CAL:DUMP? FACTory;:CAL:COPY WORKing Factory cal. Cal 4 values Cal 3 values No Change No Change No Change
6. CAL:DUMP? PRIor;:CAL:COPY WORKing Cal 4 values Factory cal. Cal 4 values No Change No Change No Change