Owner's manual
Table Of Contents
- SECTION 1 - INTRODUCTION
- SECTION 2 - INSTALLATION
- SECTION 3 - CALIBRATION
- SECTION 4 - OPERATION
- FIGURE 1-1. Remotely Controlled Power Supply Configurations Using Kepco Products
- SECTION 1 - INTRODUCTION
- 1.1 Scope of Manual
- 1.2 General Description
- TABLE 1-1. Kepco BIT 488, BIT 4882, and BIT 4886 Digital Programming Cards
- TABLE 1-2. Applicability of BIT 4882, 4882-F and BIT 4886 Cards to Specific BOP Models
- 1.3 Specifications, BIT 4882, BIT 4882-F and BIT 4886
- TABLE 1-3. Specifications, BIT 4882, BIT 4882-F AND BIT 4886
- SECTION 2 - INSTALLATION
- 2.1 Unpacking and Inspection
- 2.2 Set Start-up DefaultS
- 2.2.1 Set (GPIB) Device Address
- 2.2.2 Start-up Language Default
- 2.2.3 IEEE Cable Shield Ground Selection
- 2.2.4 Set Power Supply Identification Switch
- FIGURE 2-1. BIT 4882 and BIT 4882-F Switch and Adjustment Locations
- TABLE 2-1. Power Supply Identification Switch S2 Setting
- 2.3 Installation of Interface Card into the BOP
- TABLE 2-2. Device Address Selection
- FIGURE 2-2. Installation of Model BIT 4882-f into BOP
- FIGURE 2-3. Installation of Model BIT 4882 into BOP
- 2.4 Input/OUtput Signals
- TABLE 2-3. Input/Output Pin Assignments
- FIGURE 2-4. IEEE 488 (GPIB) Connector
- SECTION 3 - CALIBRATION
- 3.1 Equipment Required
- 3.2 Adjustment of the Bop ±10 Volt Calibration Controls (R31, R32)
- 3.3 Adjustment of the Ammeter Zero (R50)
- FIGURE 3-1. BOP Power Supply, Internal Calibration Control Locations
- 3.4 Adjustment Of The Output Voltage Zero (R81)
- 3.5 Adjustment of the Full Scale Output Voltage (R21)
- TABLE 3-1. BOP Power Supply, Internal Calibration Controls
- 3.6 Voltage Reading Zero Calibration (R35)
- 3.7 Voltage Reading Calibration (R19)
- 3.8 Adjustment of the Output Current Zero (R83)
- 3.9 Adjustment of the Full Scale Output Current (R22)
- 3.10 Current Reading Zero Calibration (R36)
- TABLE 3-2. Calibration Measurements and Tolerances - Current
- TABLE 3-3. Suggested Sense Resistors
- FIGURE 3-2. Current Shunt Connections
- 3.11 Current Reading Calibration (R20)
- SECTION 4 - OPERATION
- 4.1 General
- 4.2 IEEE 488 (GPIB) Bus Protocol
- TABLE 4-1. IEEE 488 (GPIB) Bus Interface Functions
- TABLE 4-2. IEEE 488 (GPIB) Bus Command Mode Messages
- TABLE 4-3. IEEE 488 (GPIB) Bus Data Mode Messages
- 4.3 SCPI Programming
- 4.3.1 SCPI Messages
- 4.3.2 Common Commands/Queries
- 4.3.3 SCPI Subsystem Command/Query Structure
- FIGURE 4-1. Tree Diagram of SCPI Commands Used with BIT 4882, BIT 4882-F and BIT 4886 Interface Card
- 4.3.4 Program Message Structure
- TABLE 4-4. SCPI Command Index
- 4.3.4.1 Keyword
- TABLE 4-5. Rules Governing Shortform Keywords
- FIGURE 4-2. Message Structure
- 4.3.4.2 Keyword Separator
- 4.3.4.3 Query Indicator
- 4.3.4.4 Data
- 4.3.4.5 Data Separator
- 4.3.4.6 Message Unit Separator
- 4.3.4.7 Root Specifier
- 4.3.4.8 Message Terminator
- 4.3.5 Understanding The Command Structure
- 4.3.6 Program Message Syntax Summary
- 4.3.7 SCPI Program Example
- FIGURE 4-3. Typical Example of Interface Card Program Using SCPI Commands
- 4.4 CIIL Programming
- APPENDIX A - SCPI COMMON COMMAND/QUERY DEFINITIONS
- TABLE A-1. IEEE 488.2 Command/query Index
- A.2 *CLS — Clear Status Command
- A.3 *ESE — Standard Event Status Enable Command
- TABLE A-2. Standard Event Status Enable Register and Standard Event Status Register Bits
- A.4 *ESE? — Standard Event Status Enable Query
- A.5 *ESR? — Event Status Register Query
- A.6 *IDN? — Identification Query
- A.7 *OPC — Operation Complete Command
- A.8 *OPC? — Operation Complete Query
- A.9 *RST — Reset Command
- FIGURE A-1. GPIB Commands
- A.10 *SRE — Service Request Enable Command
- TABLE A-3. Service Request Enable and Status Byte Register Bits
- A.11 *SRE? — Service Request Enable Query
- A.12 *STB? — Status Byte Register Query
- A.13 *TRG — Trigger Command
- A.14 *TST? — Self Test Query
- A.15 *WAI — Wait-To-Continue Command
- APPENDIX B - SCPI COMMAND/QUERY DEFINITIONS
- B.1 Introduction
- TABLE B-1. SCPI Subsystem Command/query Index
- B.2 INITiate[:IMMediate] Command
- B.3 INITiate:CONTinuous Command
- B.4 INITiate:CONTinuous Query
- FIGURE B-1. Programming the Output
- B.5 MEASure[:SCALar]:CURRent[:DC]? Query
- B.6 MEASure:VOLTage[:SCALar][:DC]? Query
- B.7 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] Command
- B.8 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] Query
- B.9 [SOURce:]CURRent:[:LEVel]TRIGgered[:AMPlitude] Command
- B.10 [SOURce:]CURRent:[:LEVel]TRIGgered[:AMPlitude]? Query
- B.11 .[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude] Command
- B.12 [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]? Query
- B.13 [SOURce:]VOLTage:[:LEVel]TRIGgered[:AMPlitude] Command
- B.14 [SOURce:]VOLTage:[:LEVel]TRIGgered[:AMPlitude]? Query
- B.15 [SOURce:]FUNCtion:MODE
- B.16 STATus:OPERation:CONDition Query
- TABLE B-2. Operation Condition Register, Operation Enable Register, and Operation Event Register ...
- B.17 STATus:OPEReration:ENABle Command
- B.18 STATus:OPEReration:ENABle? Query
- B.19 STATus:OPERation[:EVENt] Query
- B.20 STATus:PRESet Command
- B.21 STATus:QUEStionable[:EVENt]? Query
- TABLE B-3. Questionable Event Register, Questionable Condition Register and Questionable Conditio...
- B.22 STATus:QUEStionable:CONDition? Query
- FIGURE B-2. Using Status Commands and Queries
- B.23 STATus:QUEStionable:ENABle Command
- B.24 STATus:QUEStionable:ENABle? Query
- B.25 SYSTem:ERRor? Query
- TABLE B-4. Error Messages
- B.26 SYSTem:LANGuage Command
- B.27 SYSTem:VERSion Query
- FIGURE B-3. Using System Commands and Queries
- APPENDIX C - CIIL COMMAND DEFINITIONS
- TABLE C-1. CIIL Subsystem Command/query Index
- FIGURE C-1. FNC — Function Command
- FIGURE C-2. INX — Initiate Op Code Command
- FIGURE C-3. FTH — Fetch Command
- FIGURE C-4. SET Command
- FIGURE C-5. RST — Reset Command
- FIGURE C-6. CNF, IST — Confidence Test, Internal Self Test Commands
- TABLE C-2. CIIL Error Messages
- FIGURE C-7. STA — Status Command
- TABLE C-3. CIIL Error Handling Utility Commands
- FIGURE C-8. GAL — Go to alternate Language Command

BIT 4882 030507 B-7/(B-8Blank)
B.25 SYSTem:ERRor? QUERY
SYST:ERR?
Syntax: Short Form: SYST:ERR? Long Form: SYSTem:ERRor?
Return Value: <int_value,string>
Description: Posts error messages to the output queue. Returns the next error number followed by its corre-
sponding error message string from the instrument error queue. The error queue is a FIFO (first in first
out) buffer that stores errors as they occur. As it is read, each error is removed from the queue and the
next error message is made available. When all errors have been read, the query returns 0,"No error".
If more errors are accumulated than the queue can hold, it will overflow. The oldest errors stay in the
queue but the most recent errors are discarded. The last error in the queue will be -350,"Too many
errors." Error messages are defined in Table B-4.
B.26 SYSTem:LANGuage COMMAND SYST:LANG
Syntax: Short Form: SYST:LANG Long Form: SYSTem:LANGuage
Description: Allows CIIL command language (see Appendix C) to be used to program the power supply.
(CIIL is included to provide compatibility with earlier Kepco equipment.) Once CIIL is selected, the
CIIL command ‘GAL’ followed by the command ‘SCPI’ must be sent for the power supply to respond to
SCPI commands (with the exception of this command, SYST:LANG, which is always active).
B.27 SYSTem:VERSion QUERY SYST:VERS?
Syntax: Short Form: SYST:VERS? Long Form: SYSTem:VERSion?
Return Value: <int_value>.<int_value> (YYYY.V)
Description: Identifies SCPI Version implemented. Returns SCPI Version number:
YYYY = year, V = Revision number for specified year. (See example, Figure B-3.)
FIGURE B-3. USING SYSTEM COMMANDS AND QUERIES
TABLE B-4. ERROR MESSAGES
ERROR MESSAGE EXPLANATION
0,“No error” No error
-100,“Command error” Command and data understood, but more information included which is not
recognized.
-222,“Data [current or voltage] out of range” Value exceeds power supply rating
-240,“Hardware error” Power supply did not respond to command
SYST:VERS Returns 1998.0.
SYST:KLOC ON keypad locked, only remote control possible.
SYST:KLOC? Returns 1 indicating keypad locked.
SYST:KLOC OFF keypad unlocked, pressing LOCAL key allows Local mode
operation.
SYST:KLOC? Returns 0 indicating keypad unlocked.