Manual

BIT 232 022800
B-3
FIGURE B-4. MEASure:CURRent? and MEASure:VOLTage? QUERIES
Syntax
: Short Form: MEAS:CURR?
: MEAS:VOLT?
Long Form: MEASure:CURRent[:DC]?
MEASure:VOLTage[:DC]?
Function
: Measures actual current or voltage
Description:
This query returns the actual value of output current or voltage (measured at the sense terminals) as
determined by the commanded value of voltage and current and load conditions.
Example:
VOLT 21; CURR 5 Power supply output commanded to go to 21V, 5A
MEAS:CURR? Controller reads actual value of output current, e.g., 4.83A.
MEAS:VOLT? Controller reads actual value of output voltage, e.g., 20.9.
MEAS:CURR?
MEAS:VOLT?