Owner manual

ABC 040104
4-3
4. LCD will flash Iout F.S.adj. **WAIT** then Iout F.S. adj < >. Monitor DVM and
use or keys to increase or decrease output voltage as necessary to get DVM reading
as close as possible to value listed below (without exceeding value listed). If the
"
and
#
keys do not provide the required adjustment range, press the 1 (increase) or 3 (decrease)
key for additional range.
ABC 10-10DM: 1.000V
ABC 15-7DM: 0.700V
ABC 25-4DM: 1.000V
ABC 36-3DM: 0.750V
ABC 60-2DM: 2.000V
ABC 125-1DM: 1.000V
5. Press ENTER when complete.
4.4 CHANGING THE CALIBRATION PASSWORD
The 4-digit password is required for to enter calibration mode. The factory default passwords
are listed in Table 4-1.
1. If the unit is not in command entry status , press CLEAR or RESET to put the unit in com-
mand entry status.
2. Press MENU key until LCD reads Change Passw. OldPass .
3. Enter the old password and press ENTER. Then enter the new 4-digit Password and press
ENTER.
4. The new password is accepted and the LCD shows the next Menu screen GoTo Previous
CALIB? 1 Yes.
5. Press CLEAR to exit the Menu, press 1 to restore the previous calibration values (see PAR.
4.5), or MENU for additional menu screens.
TABLE 4-1. FACTORY DEFAULT CALIBRATION PASSWORDS
MODEL PASSWORD
ABC 10-10DM 1010
ABC 15-7DM 1507
ABC 25-4DM 2504
ABC 36-3DM 3603
ABC 60-2DM 6002
ABC 125-1DM 1251