Manual
Page 11
SETTING MAIN MENU OPTIONS
The following functions are available in the SmartSonic Main Menu:
NOTES: After you select any option the “Tank Graphic” will cycle a few times. Wait until one of the
options starts to blink before proceeding to the next step.
When you change an option, a flashing “Ar. 5” may appear. This is just to remind you that the option
you just selected may affect the Relay or 4-20mA Settings, and that you need to review these settings.
Setting the Indication Mode
The first function in the Main Menu is the Indication Mode. This sets the type of measurement. The options are:
• DIST: The displayed reading represents the distance from the sensor face to the process media.
• LEVEL: The displayed reading represents the level of the process media measured from the bottom of the vessel.
• FLOW: The displayed reading represents the flow rate through the configured Flume or Weir.
• DIA: DO NOT USE
Setting an Indication Mode enables you to toggle between other indication types as well, using the BACK and NEXT
buttons as described in the table below: For example, if the unit is configured for Level, the BACK and NEXT buttons will
toggle between Level and Distance.
Indication Toggle Selections (BACK / NEXT Buttons)
Distance Distance / Level
Level Level / Distance
Flow Flow / Level / Distance / Totalization (High) / Totalization (Low)
Totalization indicates the total accumulative flow. It is expressed in 10 digits. The first 5 digits (High) are shown as
H00000, and the second 5 digits (Low) are shown as L00000. This function will be described later in this manual
(Additional Features).
Press / Action Display Description
and
Move between the available options. DIA
(Diameter) is no longer available.
For Example
Select the desired option. This will show on the
display for a few moments, then return to the
Main Menu
Go to the next function (Measurement Mode)
Functions Page
Setting the Indication Mode 11
Setting the Measurement Units 12
Setting the Relay Values 12
Setting The 20mA and 4mA Values 17
Setting the Open Channel Flow Measurements 17
Setting the Tank Height 17
Setting the Application Type 18
Setting the Operation Modes 18
Setting the Sensor Offset 19
Setting the Scan Distance Values 20
Clearing the Scan Distance Values 21