Restoring the Default Settings
The Pr.10 program allows you to clear programs Pr.00, Pr.02 – Pr.09, and reverts to Factory Settings:
Pr.00 – GPM 1U01 or M3/Hr 1E01
Pr.02 – Sbd 00.00, E000, Tank Height = Default
Pr.03 – Resets all interfering signals
Pr.04 – Level device L000 and 00.00; Flow Device F000 and 00.00
Pr.05 – Level device Tank Height = Pr.02; Flow device 55500 M3/Hr or 244400 GPM
Pr.06 – SE 0
Pr.07 – Tank Height = Pr.02
Pr.08 – d000
Pr.09 – 01.00
The Pr.10 program does not affect the Initial Set-up Programs (Pr.20 – Pr.24)
Press / Action Display Description
and simultaneously
Enter the Program Menu
Nine Times
Cycle to Program #10
Default Screen (while entering program)
Reverts settings to Factory Settings
Returns to Pr.10
OPEN CHANNEL FLOW MEASUREMENT
This section describes how to set the flow measurement parameters for Open Channel Flow measurement. It also covers
the flume/weir codes used when selecting the flume or weir of your application.
Selecting the Open Channel Flow Settings
The Pr.00 program enables you to select one of the preset flumes/weirs for Open Channel Flow measurement. This
program is only available when Flow is selected in Pr.23 (Initial Set-up Programs).
Using Pr.00 program, the data is entered in the following format: , where:
X = Flume / Weir Style
U = U.S. Standard (U) or European (E) type flume / weir
FF = The specific flume / weir dimensions
The “X” and “FF” codes are found on the following pages, grouped by “U” U.S. Standard and “E” European types.
Refer to the CALIBRATION section for an explanation of accessing and using the SmartSonic program menu.