Operating instructions for the high pressure calibrator (HPX) 450565.
Table of contents View 3 Notes on the operating instructions 4 1. Description of the device 4 2. General safety warnings 4 3. Operating the HPX calibrator 5 4. Description of the functions 5 5. Menu navigation for calibrators 6 6. Commissioning 6 7. Maintenance / disposal 7 8.
View 2 1 10 3 4 6 5 7 11 8 1 Port for interface cable (K-104-A or K-114-A) (PC connection / RS485) 2 Device base 3 Display 4 SELECT and ENTER buttons 5 Oil chamber 6 Recirculation pipe 7 Screwed sealing plug 8 Manual booster pump 9 Pressure connection for test object, without overpressure valve (700 bar) 10 Drain valve 11 Screw compressor 9 3
Notes on the operating instructions Commissioning Scope of delivery - The operating instructions are intended for specialist workers and trained personnel. - Before each stage of work, read the relevant notes and warnings carefully, and keep to the sequence as stated. - Pay particular attention to the section on «General safety warnings». A pressure-resistant connection for the test object is required in order to use the high pressure calibrator.
safety must be followed whenever work is carried out. Internal regulations issued by the operator must be followed, even if they are not mentioned in these instructions. Never use the high pressure calibrator together with an external pressure source. Do not remove any connected components (e.g. test objects) when the high pressure calibrator is under pressure. Open the screwed sealing plug before removing parts. Do not use Teflon tape to seal the pressure connection.
SELECT button ENTER button The SELECT button positioned on the front is used to switch the device on, to select a function and to select the various pressure units. The ENTER button positioned on the front is used to activate the selected function or pressure unit on the device. You can also press the ENTER button to switch between the minimum and maximum pressure values measured thus far. 5.
is shown, and press ENTER to activate. You can now use SELECT to choose the function you want, and ENTER to execute the function. Depending on the current setting, the first function to be shown is either min/ max disp or LEAK disp. Leak measurement function Leak mode is used to determine the pressure change over a defined period, which can be adjusted. The unit to be tested must be connected to the high pressure calibrator on the pressure side. Start leak measurement Activate the MANO menu.
Disposal This product must not be disposed of as normal household waste at the end of its useful lifetime. To prevent possible damage to the environment or to health due to uncontrolled waste disposal, this product must be separated from other waste and recycled correctly in order to ensure sustainable use of the raw materials. Control Center CCS30 8. Software for calibrators The USB interface converter (K-114-A) enables communication between the calibrator and a computer.
Spare parts and accessories for KELLER pressure calibrators suitable for Description Article number Carrying case, empty LPX MPX HPX 309025.0005 x x x Battery, type CR2430 557005.0001 x x x Hose nipple 508832.0005 x x Connecting nipple, G 1/4” including sealing ring 508832.0004 x x Y-coupler 307025.0001 Test object adapter, G 1/4”M-G 3/8”F 506810.0028 x Test object adapter, G 1/4”M-G 1/2”F 506810.0013 x Sealing ring, G 1/8” 508635.0001 x x Sealing ring, G 1/4” 508635.
EU-Konformitätserklärung EU Declaration of Conformity Déclaration UE de Conformité Herewith we declare, that the follow- Nous attestons que le produit ou ing product or product range gamme de produits : Hochdruckkalibrator HPX High pressure calibrator HPX Calibrateur haute pression HPX wird hiermit bestätigt, dass es den An- meet the basic requirements, which are répondent aux exigences prévues forderungen folgender EU-Richtlinien established in the guidelines of the par les directives de la Co