User Manual

Model 2470
High Voltage SourceMeter Instrument User's Manual Section 6: Measure I-
V characteristics of FETs
2470-900-01 Rev. A / May 2019 6-7
Subthreshold current or drain family of curves measurement
device connections
Both a subthreshold current and a drain family of curves measurement use the same instrument
connections. Both 2470s are configured to source voltage and measure current.
In this circuit:
Connect the Force HI terminal of 2470 #1 to the gate of the FET.
Connect the Force HI terminal of 2470 #2 to the drain of the FET.
Connect the source of the FET to the Force LO terminals of both 2470s.
To source and measure from all three terminals of the FET, you need a third 2470.
The following figure shows an I-V test configuration for a FET using two 2470s.
Figure 33: Three-terminal I-V test configuration for a FET