High Voltage SourceMeter Instrument User's Manual Section 5:
Leakage current and insulation resistance
2470-900-01 Rev. A / May 2019 5-9
The graph in the following figure shows the results of this application. Notice the exponential current
response of the capacitor as it charges up to 300 V over time.
Figure 25: Leakage current test results
Measuring insulation resistance
The following application demonstrates how to use the 2470 to measure insulator resistance. The
application shows how to use the front-panel interface, the remote interface using the SCPI command
set, and the remote interface using the Test Script Processor (TSP
®
) command set.
You typically make insulation resistance measurements between traces on a printed circuit board and
between conductors in cables and connectors.
This application sets the 2470 to source 700 V and makes 10 resistance readings with a 100 ms
interval time. Once the measurements are made, the output is turned off.
For this test, you will:
Reset the instrument
Set the instrument to read the rear terminals
Select the source voltage function and measure resistance function
Set the magnitude of the voltage source output
Turn on autoranging
Use the Simple Loop trigger model template to set the numbers of readings and interval time
Make readings