Datasheet

Datasheet
TEK.COM10
General Characteristics
(Default mode unless specified)
Factory Default Standard Power-Up Setting
SCPI mode
Source Output Modes Fixed DC level
Memory/configuration list (mixed function)
Stair (linear and logarithmic)
Memory Buffer >5,000,000 readings with selected measured values and timestamp
Real-Time Clock Lithium battery backup (more than 3 years of battery life)
Remote Interfaces GPIB: IEEE Std 488.1 compliant; supports IEEE Std 488.2 common commands and status model topology
USB device (rear panel, type B): 2.0 full-speed USBTMC
USB host (front panel, type A): USB 2.0, support for flash drives, FAT32
Ethernet: RJ-45 connector, 10/100 BT
IP Configuration Static or DHCP
Expansion Interface The TSP-Link
®
expansion interface allows TSP-enabled instruments to trigger and communicate with
eachother
LXI Compliance LXI version 1.4 Core 2011
TSP Mode Embedded Test Script Processor (TSP) accessible from any host interface
Display Five-inch capacitive touch, color TFT WVGA (800 × 480) with LED backlight
Input Signal Connections Front: Banana. Rear: High-voltage triaxial
Programmability SCPI or TSP command sets
Interlock Active high-input
Digital I/O Lines: Six input/output, user-defined, for digital I/O or triggering
Connector: 9-pin female D
Input Signal Levels: 0.7 V (maximum logic low), 3.7 V (minimum logic high)
Input Voltage Limits: –0.25 V (absolute minimum), +5.25 V (absolute maximum)
Maximum Source Current: +2.0 mA at > 2.7 V (per pin)
Maximum Sink Current: –50 mA at 0.7 V (per pin, solid-state fuse protected)
5 V Power Supply Pin: Limited to 500 mA at > 4 V (solid-state fuse protected)
Handler: User-definable start of test, end of test, four category bits
Cooling Forced air, variable speed
Overtemperature Protection Internally sensed temperature overload puts instrument in standby mode
Power Supply 100 VRMS to 240 VRMS, 50 Hz or 60 Hz (automatically detected at power up)
VA Rating 220 VA maximum
Altitude Maximum 2000 meters (6562 feet) above sea level
EMC Conforms to European Union EMC Directive
Safety NRTL listed to UL61010-1 and UL61010-2-30; conforms to European Union Low Voltage Directive
RoHS Conforms to European Union Restriction on Hazardous Substances Directive
Vibration MIL-PRF-28800F Class 3 Random
Warm up One hour to rated accuracies