Section 10: Capturing high-power pulse waveforms Model 2461 Interactive SourceMeter® Instrument User's Manual
10-6 2461-900-01 Rev. A / November 2015
The following figure illustrates these test parameters.
Figure 59: Timing diagram for the example
The trigger model of the pulse command
The built-in pulse settings screen and the remote pulse commands configure the trigger model of the
Model 2461 to generate the pulses. Executing the pulse command or generating a pulse setup from
the front panel takes a snapshot of the active source and digitize settings of the Model 2461. These
settings are saved to configuration lists. When you initiate the trigger model, the instrument recalls the
settings in the configuration lists, triggers the pulse and digitize actions using constant delays, and
loops to create the number of specified pulses.
The trigger model must be generated again if source or digitize settings (such as Sample Rate and
Source Readback) are changed and you want these settings to be applied to the pulse trigger model.