Section 5: Measuring low-resistance devices Model 2461 Interactive SourceMeter® Instrument User's Manual
5-6 2461-900-01 Rev. A / November 2015
View the measurements on the front-panel GRAPH swipe screen
You can view the resistance measurements as a function of time on the front-panel GRAPH swipe
screen. To access the GRAPH swipe screen, swipe the bottom part of the Home screen to the right.
A graph similar to the one in the figure below is displayed.
Figure 24: GRAPH swipe screen
To see the graph on the full screen, touch the graph icon on the GRAPH swipe bar to open the Graph
screen.
View the buffer statistics on the front panel
You can view the measurement statistics on the Model 2461 front-panel STATISTICS swipe screen,
including:
Peak-to-peak value
Minimum, maximum, and average reading values
Standard deviation
Figure 25: STATISTICS swipe screen
Set up the low-resistance application using SCPI commands
The following sequence of SCPI commands makes 100 low-resistance measurements by sourcing
current and measuring resistance. In this example, the source current magnitude and limit voltage are
set automatically. It uses remote commands to change the front-panel display to show the GRAPH
swipe screen. This allows you to view numeric data at the top of the screen and graphic data at the
bottom of the screen.