Datasheet

Model 2461 Interactive SourceMeter® Instrument User's Manual Section 2: Using the front-panel interface
2461-900-01 Rev. A / November 2015 2-5
When the source of the instrument is turned off, it may not completely isolate the instrument from the
external circuit. You can use the Output Off setting to place the Model 2461 in a known,
noninteractive state during idle periods, such as when you are changing the device under test. The
output-off states that can be selected for a Model 2461 are normal, high-impedance, zero, or guard.
See "Output-off state" in the Model 2461 Reference Manual for additional details.
Using the front panel:
Press the OUTPUT ON/OFF switch. The instrument is in the output-on state when the switch is
illuminated. The instrument is in the output-off state when the switch is not illuminated.
Using SCPI commands:
To turn the output on, send the command:
:OUTPut:STATe ON
To turn the output off, send the command:
:OUTPut:STATe OFF
Using TSP commands:
To turn the output on, send the command:
smu.source.output = smu.ON
To turn the output off, send the command:
smu.source.output = smu.OFF
Touchscreen display
The touchscreen display gives you quick front-panel access to source and measure settings, system
configuration, instrument and test status, reading buffer information, and other instrument
functionality. The display has multiple swipe screens that you can access by swiping the front panel.
You can access additional interactive screens by pressing the front-panel MENU, QUICKSET, and
FUNCTION keys.
Do not use sharp metal objects, such as tweezers or screwdrivers, or pointed objects, such as pens
or pencils, to touch the touchscreen. It is strongly recommended that you use only fingers to operate
the instrument. Use of clean-room gloves to operate the touchscreen is supported.
Select items on the touchscreen
To select an item on the displayed screen, do one of the following:
Touch it with your finger
Turn the navigation control to highlight the item, and then press the navigation control to select it
The following topics describe the Model 2461 touchscreen in more detail.