Datasheet
DIGITAL MULTIMETERS & SYSTEMS
www.keithley.com
1.888.KEITHLEY
(U.S. only)
A Greater Measure of Confidence
A Tektronix Company
DMM7510
7½-Digit Graphical Sampling Multimeter
SIGNAL CHARACTERISTICS
61, 62, 63
TYPICAL AC AND DC COUPLED
Range
Analog
Bandwidth
(–3dB)
Maximum Flatness
Error
3 Hz to 20 kHz
64
THD
20 kHz Signal
(–1dB FS)
65
DC-coupled
Settling Time
(0.5%)
AC-coupled Filter
FAST Settling Time
(0.5%)
AC-coupled Filter
SLOW Settling Time
(0.5%)
AC Coupling
Low Frequency
(–3dB) point
66
100.000 mV 600 kHz 0.015 dB 0.04 % 5 µs 80 ms 2.3 s 1 Hz
1.00000 V 600 kHz 0.01 dB 0.03 % 6 µs 80 ms 2.5 s 1 Hz
10.0000 V 600 kHz 0.01 dB 0.01 % 4 µs 80 ms 2.5 s 1 Hz
TYPICAL DC COUPLED
Range
Analog
Bandwidth
(–3dB)
Maximum Flatness
Error
3 Hz to 1 kHz
64
Total Harmonic
Distortion (THD)
1 kHz Signal
(–1dB FS)
65
Settling Time
(0.5%)
100.000 V 20 kHz
67
0.1 dB 1.3 % 160 µs
1000.00 V 20 kHz 0.1 dB 1.8 % 80 µs
TYPICAL AC COUPLED
Range
Analog
Bandwidth
(–3dB)
Maximum Flatness
Error
3 Hz to 20 kHz
64
Filter FAST
Settling Time
(0.5%)
Filter SLOW
Settling Time
(0.5%)
Low Frequency
Coupling Point
66
(–3dB)
100.000 V 600 kHz 0.1 dB 80 ms 2.3 s 1 Hz
1000.00 V 600 kHz 0.1 dB 80 ms 2.3 s 1 Hz
DC-COUPLED ADDITIONAL NOISE UNCERTAINTY, TYPICAL
68
DC-COUPLED TOTAL HARMONIC DISTORTION (THD), TYPICAL
69
0.1
1
10
100
1000
1k 2k 5k 10k 20k 50k 100k 200k 500k 1M
ppm of Range
Samples per Second
RMS Noise, Digitize Voltage
100mV
1V
10V
100V
1000V
-120
-110
-100
-90
-80
-70
-60
-50
1k 2k 5k 10k 20k 50k 100k 200k
dB of Range
Samples per Second
Total Harmonic Distortion, Digitize Voltage
100mV
1V
10V
100V
1000V
61. Accuracy with sample rate 1M per second and aperture 1µs.
62. Verified with sine wave input and DC content ≤3% of range.
63. For AC coupling, maximum crest factor of 5.
64. For DC coupled, 0dB reference frequency is 3Hz. For AC coupled, 0dB reference frequency is
1kHz. For AC coupled operation below 1kHz, add 0.1dB.
65. Exclusive of source input noise.
66. With AC coupling frequency = 3Hz and AC coupling filter = Slow.
67. For input impedance auto, bandwidth is 6kHz.
68. Specified with aperture auto and 4-wire short on input terminals. For 100V range, input
impedance 10MΩ, multiply by 2.5. For all ranges and sample rate >1k, add an additional 3×
RMS noise uncertainty to ppm of range.
69. Specified with aperture Auto, 100 Hz sine wave for sample rate ≤ 5 k, and 1 kHz sine wave for
sample rate ≥ 10 k. Distortion is calculated using first five harmonics.
Model DMM7510 condensed specifications
Model DMM7510 condensed specifications