Datasheet
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications are subject to change without notice.
SPEC-DMM7510 Rev. B / October 2016 Page 9 of 25
RMS NOISE (ADDITIONAL PEAK NOISE UNCERTAINTY)
42
Applies to ± ppm of range
Peak noise uncertainty is included in DC Specifications for PLC ≥ 1
Add peak noise uncertainty to measurements for PLC < 1
Examples:
1 mA at 0.006 PLC: 9 (from Accuracy table) + 20 (additional peak noise uncertainty) = 29 ppm of range
1 mA at 1 PLC: 9 + 0 = 9 ppm of range
NPLC Digits 10 µA 100 µA 1 mA 10 mA 100 mA 1 A 3 A 10 A
43
5 7½ 0.15 0.14 0.09 0.1 0.3 0.3 0.2 0.8
1 7½ 0.4 0.13 0.1 0.1 0.5 0.5 0.3 1.2
0.2
44
6½ 0 (220) 0 (23) 0.2 (3.4) 0.2 (1.6) 2 (10) 2 (11) 0.7 (4.6) 4 (32)
0.2 6½ 120 (260) 12 (26) 1.2 (3.8) 0.3 (1.8) 1.9 (9.8) 2 (10) 0.8 (5) 8 (37)
0.06 5½ 130 (280) 12 (29) 1.3 (5.6) 0.4 (3.9) 2 (14) 2 (14) 1.2 (7.7) 10 (59)
0.006 4½ 130 (350) 14 (42) 3 (20) 2 (20) 4 (30) 4 (31) 7 (51) 20 (110)
0.0005 3½ 260 (2110) 30 (300) 20 (150) 20 (160) 30 (190) 30 (190) 70 (510) 60 (420)
DC CURRENT CHARACTERISTICS
Range 10 µA 100 µA 1 mA 10 mA 100 mA 1 A 3 A 10 A
43
Effective internal shunt value
45
1 kΩ 100 Ω 10 Ω 1 Ω 0.1 Ω 0.1 Ω 0.1 Ω 0.005 Ω
Autozero off error: For ±1 °C and ≤ 10
minutes add ±(8 ppm of reading + range
error)
150 pA 1.5 nA 15 nA 150 nA 15 µA 150 µA 150 µA 3 mA
Overload recovery:
For each additional sustained amp beyond
±1.5 A, add the following initial ppm of range
error until thermally settled after overload
recovery
15500 1800 150 150 6500 200
42
Noise values are based on 1000 readings with autozero on and AMPS terminal open. RMS noise is typical. Additional peak noise is
guaranteed.
43
Rear input terminals only.
44
With line sync on.
45
Values are typical and guaranteed by design.