Datasheet
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications and characteristics are subject to change without notice.
Page 8 of 25 SPEC-DMM7510 Rev. B / October 2016
DC CURRENT
ENHANCED ACCURACY (WITHIN 30 DAYS OF AUTOCALIBRATION, T
OPER
±5 °C FROM T
ACAL
)
Range
38
Resolution
Maximum
burden
voltage
Accuracy ±[ppm of reading + ppm of range]
24 hour
T
CAL
±1 °C
39
90 day
T
CAL
±5 °C
1 year
T
CAL
±5 °C
2 year
T
CAL
±5 °C
Temperature
coefficient
40
10.000000 µA 1 pA 15 mV 30 + 30 75 + 30 75 + 30 75 + 30 0.15 + 0.1
100.00000 µA 10 pA 15 mV 20 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0.1
1.0000000 mA 100 pA 15 mV 30 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0.1
10.000000 mA 1 nA 20 mV 40 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0.1
100.00000 mA 10 nA 200 mV 50 + 18 150 + 30 150 + 30 150 + 30 0.15 + 0.1
1.0000000 A 100 nA 400 mV 150 + 50 400 + 50 400 + 50 400 + 50 0.15 + 0.1
3.000000 A 1 µA 1300 mV 200 + 40 400 + 40 400 + 40 400 + 40 0.15 + 0.1
10.000000 A
41
1 µA 650 mV 700 + 275 800 + 275 1500 + 275 2000 + 275 50 + 10
ACCURACY
Range
38
Resolution
Maximum
burden
voltage
Accuracy ±[ppm of reading + ppm of range]
24 hour
T
CAL
±1 °C
39
90 day
T
CAL
±5 °C
1 year
T
CAL
±5 °C
2 year
T
CAL
±5 °C
Temperature
coefficient
40
10.000000 µA 1 pA 15 mV 30 + 30 100 + 30 125 + 40 175 + 50 10 + 8
100.00000 µA 10 pA 15 mV 20 + 5 75 + 12 100 + 15 150 + 20 10 + 3
1.0000000
mA
100 pA 15 mV 30 + 5 75 + 12 100 + 15 150 + 20 10 + 3
10.000000
mA
1 nA 20 mV 40 + 5 75 + 12 100 + 15 150 + 20 10 + 3
100.00000
mA
10 nA 200 mV 50 + 18 300 + 30 400 + 30 500 + 30 50 + 5
1.0000000 A 100 nA 400 mV 150 + 50 400 + 50 450 + 50 500 + 50 10 + 10
3.000000 A 1 µA 1300 mV 200 + 40 400 + 40 450 + 40 500 + 40 10 + 10
10.000000 A
41
1 µA 650 mV 700 + 275 800 + 275 1500 + 275 2000 + 275 50 + 10
38
20 % overrange supported for all ranges except for 3 A and 10 A, which are 1 % supported.
39
Relative to calibration accuracy.
40
Add per degree from T
CAL
±5
o
C.
41
Rear input terminals only.