Datasheet

Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications are subject to change without notice.
SPEC-DMM7510 Rev. B / October 2016 Page 7 of 25
RMS NOISE (ADDITIONAL PEAK NOISE UNCERTAINTY)
36
Applies to ± ppm of range
Peak noise uncertainty is included in DC specifications for ≥ 1 PLC
Add peak noise uncertainty to measurements when < 1 PLC
Examples:
10 Ω at 0.2 PLC: 80 (from Accuracy table) + 230 (additional peak noise uncertainty) = 310 ppm of range
10 Ω at 1 PLC: 80 + 0 = 80 ppm of range
NPLC Digits
1
10
100
1 k
10 k
5 10 11 6 5 0.9
1 9 9 7 7 0.8
0.2
37
30 (130) 30 (120) 30 (120) 30 (120) 3 (16)
0.2 60 (220) 60 (230) 50 (190) 50 (190) 9 (35)
0.06 70 (350) 70 (350) 50 (290) 50 (280) 20 (90)
0.006 130 (750) 120 (830) 110 (700) 100 (690) 20 (110)
0.0005 520 (3550) 530 (3520) 530 (3380) 500 (3370) 100 (670)
DRY CIRCUIT RESISTANCE CHARACTERISTICS
Maximum 4-wire ohm lead resistance
0.5 per lead for 1 range
10 % of range per lead for 10 to 100 ranges
50 per lead for 1 k to 10 k ranges
Input current limit
For signals > [±20 mV], current limited, ±13 mA, typical
Offset compensation
Selectable on 1 to 10 k ranges
Autozero off error
For ±1 °C and ≤ 10 minutes, add ± 8 ppm of reading
36
Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. RMS noise is typical. Additional peak noise is
guaranteed.
37
With line sync on.