Datasheet

Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications and characteristics are subject to change without notice.
Page 6 of 25 SPEC-DMM7510 Rev. B / October 2016
RESISTANCE CHARACTERISTICS
Maximum 4-wire ohms lead resistance
5 per lead for 1 range, 10 % of range per lead for 10 to 1 k
ranges; 1 k per lead for all other ranges
Offset compensation
Selectable on 4-wire, 1 to 100 k ranges
Open lead detector
Default is off
Autozero off error
For 2-wire ohms, ±1 °C and ≤ 10 minutes, add ±(8 ppm of reading)
and 1.5 m for 10 range, 15 m for 100 and 1 k ranges,
150 m for 10 k range, 1.5 for 100 k range, and 15 for all
other ranges
For 4-wire ohms, ±1 °C and ≤ 10 minutes, add ±(8 ppm of reading)
Input current limit
For signals with a magnitude of +12 V to +40 V or -12 V to -40 V:
±13 mA source or sink, typical
For signals with a magnitude of greater than +40 V or -40 V:
±130 µA source or sink, typical
DRY CIRCUIT RESISTANCE
ENHANCED ACCURACY (WITHIN 30 DAYS OF AUTOCALIBRATION, T
OPER
±5 ºC FROM T
ACAL
)
Range
31
Resolution Test
current
35
(±5 %)
Open circuit
DUT
voltage
32
Accuracy ±[ppm of reading + ppm of range]
24 hour
T
CAL
±1 °C
33
90 day
T
CAL
±5 °C
1 year
T
CAL
±5 °C
2 year
T
CAL
±5 °C
Temperature
coefficient
34
1.000000 1 µ
10 mA 25 mV 25 + 80 50 + 80 50 + 80 50 + 80 1.5 + 0.1
10.00000 10 µ
1 mA 25 mV 25 + 80 50 + 80 50 + 80 50 + 80 1.5 + 0.1
100.0000 100 µ
100 µA 25 mV 25 + 80 90 + 80 90 + 80 90 + 80 1.5 + 0.1
1.000000 k 1 m
10 µA 25 mV 25 + 80 180 + 80 180 + 80 180 + 80 1.5 + 0.1
10.00000 k 10 m
5 µA 25 mV 25 + 80 320 + 80 320 + 80 320 + 80 1.5 + 0.1
ACCURACY
Range
31
Resolution Test
current
35
(±5 %)
Open circuit
DUT
voltage
32
Accuracy ±[ppm of reading + ppm of range]
24 hour
T
CAL
±1
°C
33
90 day
T
CAL
±5 °C
1 year
T
CAL
±5 °C
2 year
T
CAL
±5 °C
Temperature
coefficient
34
1.000000 1 µ
10 mA 25 mV 25 + 80 50 + 80 70 + 80 90 + 80 2.5 + 1
10.00000 10 µ
1 mA 25 mV 25 + 80 50 + 80 70 + 80 90 + 80 5 + 1
100.0000 100 µ
100 µA 25 mV 25 + 80 90 + 80 140 + 80 200 + 80 2.5 + 1
1.000000 k 1 m
10 µA 25 mV 25 + 80 180 + 80 400 + 80 600 + 80 5 + 1
10.00000 k 10 m
5 µA 25 mV 25 + 80 320 + 80 800 + 80 1300 + 80 8 + 1
31
20 % overrange on all ranges, except 2.4 k for the 10 K range.
32
Maximum clamp voltages are DC, typical accuracy is ±20 %. Add 20 % for offset compensation on.
33
Relative to calibration accuracy.
34
Add per degree from T
CAL
±5
o
C.
35
Test current with offset compensation off.