Datasheet

Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications are subject to change without notice.
SPEC-DMM7510 Rev. B / October 2016 Page 5 of 25
4-WIRE OHMS (≤ 10 k) OFFSET COMPENSATION ON
RMS NOISE (ADDITIONAL PEAK NOISE UNCERTAINTY)
29
Applies to ± ppm of range
Peak noise uncertainty is included in DC specifications for ≥ 1 PLC
Add peak noise uncertainty to measurements for < 1 PLC
Examples
1 k at 0.006 PLC: 3 (from Accuracy table) + 26 (additional peak noise uncertainty) = 29 ppm of range
1 k
at 1 PLC: 3 + 0 = 3 ppm of range
NPLC
Digits
1
10
100
1 k
10 k
5 2.8 0.3 0.3 0.07 0.3
1 4.2 0.4 0.4 0.12 0.5
0.2
30
30 (160) 3 (13) 3 (13) 0.4 (2.6) 1.2 (8.2)
0.2 50 (250) 5 (22) 5 (22) 0.6 (3.2) 1.2 (8.3)
0.06 110 (490) 11 (47) 11 (46) 1.1 (6.6) 2 (16)
0.006 110 (710) 10 (70) 10 (70) 4 (26) 10 (60)
0.0005 520 (3420) 50 (340) 50 (340) 40 (220) 50 (300)
2-WIRE OHMS
RMS NOISE (ADDITIONAL PEAK NOISE UNCERTAINTY)
29
Applies to ± ppm of range
Peak noise uncertainty is included in DC specifications for ≥ 1 PLC
Add peak noise uncertainty to measurements for < 1 PLC
Examples:
10 k at 0.006 PLC: 3 (from Accuracy table) + 5 (50 mΩ with Rel ) + 43 (additional peak noise uncertainty) = 51 ppm
of range
10 k
at 1 PLC: 3 + 5 + 0 = 8 ppm of range
NPLC
Digits
10
100
1 k
10 k
5 1.1 0.8 0.1 0.2
1 0.6 0.6 0.09 0.4
0.2
30
2 (17) 2 (10) 0.2 (1.5) 0.8 (6.3)
0.2 2 (17) 2 (14) 0.3 (1.6) 0.8 (6.4)
0.06 3 (22) 3 (19) 0.4 (3.7) 2 (12)
0.006 6 (50) 6 (50) 3 (21) 6 (43)
0.0005 30 (300) 30 (230) 20 (150) 30 (210)
29
Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. RMS noise is typical. Additional peak noise is
guaranteed.
30
With line sync on.