Datasheet
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications and characteristics are subject to change without notice.
Page 4 of 25 SPEC-DMM7510 Rev. B / October 2016
ACCURACY
19
Range
20
Resolution Test current
21
(±5 %)
Accuracy ±[ppm of reading + ppm of range]
24 hour
TC
AL
±1 °C
22
90 day
T
CAL
±5 °C
1 year
T
CAL
±5 °C
2 year
T
CAL
±5 °C
Temperature
coefficient
23
1 Ω 0.1 µΩ
10 mA 15 + 50 40 + 50 50 + 50 70 + 50 2.5 + 5
10 Ω 1 µΩ
10 mA 15 + 5 40 + 5 50 + 5 70 + 5 2.5 + 0.5
100 Ω 10 µΩ
1 mA 12 + 4 35 + 4 47 + 4 65 + 4 5 + 0.25
1 kΩ 100 µΩ
1 mA 12 + 3 30 + 3 41 + 3 65 + 3 5 + 0.25
10 kΩ
24
1 mΩ
100 µA 10 + 3 30 + 3 42 + 3 65 + 3 2.5 + 0.25
100 kΩ
24,25
10 mΩ
10 µA 13 + 3 38 + 3 50 + 3 65 + 3
5 + 1
1 MΩ
24,26
100 mΩ
10 µA 14 + 3 38 + 5 50 + 5 65 + 5
5 + 1
10 MΩ
27
1 Ω 0.69 µA ǁ 10 MΩ
150 + 6 200 + 10 400 + 10 600 + 12 70 + 1
100 MΩ
27
10 Ω 0.69 µA ǁ 10 MΩ
800 + 30 2000 + 30 2000 + 30 2600 + 30 385 + 1
1 GΩ
27
100 Ω 0.69 µA ǁ 10 MΩ
9000 + 200 9000 + 200 13000 + 200 14000 + 200 3000 + 1
RESISTANCE OPEN CIRCUIT DC VOLTAGE
28
Range
20
2-wire
Offset compensation off Offset compensation on
4-wire 4-wire
1 Ω
–
9.2 V 9.5 V
10 Ω
9.2 V
9.2 V 9.5 V
100 Ω, 1 kΩ
14.0 V
14.2 V 14.3 V
10 kΩ
9.5 V 9.5 V 0.0 V
100 kΩ, 1 MΩ
12.7 V 14.3 V
0.0 V (100 kΩ range only)
10 MΩ to 1 GΩ
6.9 V 6.9 V –
19
Specifications are for 4-wire resistance, offset compensation on for ≤10 kΩ measurements, and offset compensation off for ≥10 kΩ
measurements. 1 Ω range is 4-wire only. For 2-wire, with Rel, add 50 mΩ to ppm of range uncertainty. Without Rel and with Model 1756 test
leads, add 100 mΩ to ppm of range uncertainty.
20
20 % overrange on all ranges.
21
Test current with offset compensation off.
22
Relative to calibration accuracy.
23
Add per degree from T
CAL
±5 °C.
24
Specifications are for external cable and load capacitance < 1 nF.
25
For offset compensation on, add 10 ppm of uncertainty to ppm of reading.
26
For 4-wire, 1 MΩ, open lead detection on, add 10 ppm uncertainty to ppm of reading.
27
Specified for < 10 % lead resistance mismatch in HI and LO.
28
Open circuit voltage is typical, measured from input HI to LO, SHI and SLO open. For 1 Ω to 1 MΩ ranges using an external digital multimeter
(DMM) set to 10 MΩ input impedance; for 10 MΩ to 1 GΩ ranges, set external DMM to >10 GΩ input impedance.