Datasheet

Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications are subject to change without notice.
SPEC-DMM7510 Rev. B / October 2016 Page 3 of 25
NORMAL MODE REJECTION
For DC voltage, line frequency ±0.1 %
5 PLC 1 PLC 0.2 PLC 0.01 PLC
Line sync on
110 dB 90 dB 45 dB
Line sync off
60 dB 60 dB
COMMON MODE REJECTION
For DC voltage and 1 k unbalanced in LO terminal; AC CMRR is 70 dB
NPLC 5 1 0.2 0.2
Line sync
On On On Off
CMRR
140 dB 140 dB 120 dB 80 dB
RESISTANCE
ENHANCED ACCURACY (WITHIN 30 DAYS OF AUTOCALIBRATION, T
OPER
± 5 °C FROM T
ACAL
)
10
Range
11
Resolution Test current
12
5 %)
Accuracy ±[ppm of reading + ppm of range]
24 hour
T
CAL
±1 °C
13
90 day
T
CAL
±5 °C
1 year
T
CAL
±5 °C
2 year
T
CAL
±5 °C
Temperature
coefficient
14
1.0000000 0.1 µ
10 mA 15 + 50 30 + 50 30 + 50 30 + 50 0.15 + 0.1
10.000000 1 µ
10 mA 15 + 5 30 + 5 30 + 5 30 + 5 0.15 + 0.1
100.00000 10 µ
1 mA 12 + 4 27 + 4 27 + 4 27 + 4
0.15 + 0.1
1.0000000 k 100 µ
1 mA 12 + 3 24 + 3 24 + 3 24 + 3
0.15 + 0.1
10.000000 k
15
1 m
100 µA 13 + 3 30 + 3 30 + 3 30 + 3
0.15 + 0.1
100.00000 k
15,16
10 m
10 µA 13 + 3 30 + 3 30 + 3 30 + 3
0.15 + 0.1
1.0000000 M
15,17
100 m
10 µA 14 + 3 30 + 4 30 + 4 30 + 4
0.15 + 0.1
10.000000 M
18
1 0.69 µA ǁ 10 M
150 + 6 200 + 10 200 + 10 200 + 10 70 + 1
100.00000 M
18
10 0.69 µA ǁ 10 M
800 + 30 2000 + 30 2000 + 30 2000 + 30 385 + 1
1.0000000 G
18
100 0.69 µA ǁ 10 M
9000 + 100 9000 + 100 9000 + 100 9000 + 100 3000 + 1
10
Specifications are for 4-wire resistance, offset compensation on for ≤10 k measurements, and offset compensation off for ≥10 k
measurements. 1 range is 4-wire only. For 2-wire, with Rel, add 50 m to ppm of range uncertainty. Without Rel and with Model 1756 test
leads, add 100 m to ppm of range uncertainty.
11
20 % overrange on all ranges.
12
Test current with offset compensation off.
13
Relative to calibration accuracy.
14
Add per degree from T
CAL
±5
o
C.
15
Specifications are for external cable and load capacitance < 1 nF.
16
For offset compensation on, add 10 ppm uncertainty to ppm of reading.
17
For 4-wire 1 M, open lead detector on, add 10 ppm uncertainty to ppm of reading.
18
Specified for < 10 % lead resistance mismatch in HI and LO.