NORMAL MODE REJECTION
For DC voltage, line frequency ±0.1 %
5 PLC 1 PLC ≤ 0.2 PLC ≤ 0.01 PLC
Line sync on
110 dB 90 dB 45 dB
Line sync off
60 dB 60 dB
COMMON MODE REJECTION
For DC voltage and 1 kΩ unbalanced in LO terminal; AC CMRR is 70 dB
NPLC 5 1 0.2 ≤ 0.2
Line sync
On On On Off
CMRR
140 dB 140 dB 120 dB 80 dB
RESISTANCE
ENHANCED ACCURACY (WITHIN 30 DAYS OF AUTOCALIBRATION, T
OPER
± 5 °C FROM T
ACAL
)
10
Range
11
Resolution Test current
12
(± 5 %)
Accuracy ±[ppm of reading + ppm of range]
24 hour
T
CAL
±1 °C
13
90 day
T
CAL
±5 °C
1 year
T
CAL
±5 °C
2 year
T
CAL
±5 °C
Temperature
coefficient
14
1.0000000 Ω 0.1 µΩ
10 mA 15 + 50 30 + 50 30 + 50 30 + 50 0.15 + 0.1
10.000000 Ω 1 µΩ
10 mA 15 + 5 30 + 5 30 + 5 30 + 5 0.15 + 0.1
100.00000 Ω 10 µΩ
1 mA 12 + 4 27 + 4 27 + 4 27 + 4
0.15 + 0.1
1.0000000 kΩ 100 µΩ
1 mA 12 + 3 24 + 3 24 + 3 24 + 3
0.15 + 0.1
10.000000 kΩ
15
1 mΩ
100 µA 13 + 3 30 + 3 30 + 3 30 + 3
0.15 + 0.1
100.00000 kΩ
15,16
10 mΩ
10 µA 13 + 3 30 + 3 30 + 3 30 + 3
0.15 + 0.1
1.0000000 MΩ
15,17
100 mΩ
10 µA 14 + 3 30 + 4 30 + 4 30 + 4
0.15 + 0.1
10.000000 MΩ
18
1 Ω 0.69 µA ǁ 10 MΩ
150 + 6 200 + 10 200 + 10 200 + 10 70 + 1
100.00000 MΩ
18
10 Ω 0.69 µA ǁ 10 MΩ
800 + 30 2000 + 30 2000 + 30 2000 + 30 385 + 1
1.0000000 GΩ
18
100 Ω 0.69 µA ǁ 10 MΩ
9000 + 100 9000 + 100 9000 + 100 9000 + 100 3000 + 1
10
Specifications are for 4-wire resistance, offset compensation on for ≤10 kΩ measurements, and offset compensation off for ≥10 kΩ
measurements. 1 Ω range is 4-wire only. For 2-wire, with Rel, add 50 mΩ to ppm of range uncertainty. Without Rel and with Model 1756 test
leads, add 100 mΩ to ppm of range uncertainty.
11
20 % overrange on all ranges.
12
Test current with offset compensation off.
13
Relative to calibration accuracy.
14
Add per degree from T
CAL
±5
o
C.
15
Specifications are for external cable and load capacitance < 1 nF.
16
For offset compensation on, add 10 ppm uncertainty to ppm of reading.
17
For 4-wire 1 MΩ, open lead detector on, add 10 ppm uncertainty to ppm of reading.
18
Specified for < 10 % lead resistance mismatch in HI and LO.