Datasheet

Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications are subject to change without notice.
SPEC-DMM7510 Rev. B / October 2016 Page 25 of 25
Digital I/O Connector
9-pin female D
5 V power supply pin
Limited to 500 mA at > 4 V
(solid-state fuse protected)
Lines
Six input/output, user-defined, for
digital I/O or triggering
Input signal levels
0.7 V (maximum logic low)
3.7 V (minimum logic high)
Input voltage limits
-0.25 V (absolute minimum)
+5.25 V (absolute maximum)
Maximum source current
+2.0 mA at > 2.7 V (per pin)
Maximum sink current
-50 mA at 0.7 V (per pin, solid-state
fuse protected)
Handler
User-defined start of test, end of test,
four category bits
Math functions
Rel, dB, Limit Test, Percentage, 1/x, and mX + b
Remote interface
LAN: RJ-45 connector, 10/100BT; Virtual Front Panel
GPIB: IEEE-488.1 compliant. Supports IEEE-488.2 common
commands and status model topology
USB device (rear panel, type B): 2.0 full speed, USBTMC compliant
USB host (front panel, type A): USB 2.0, support for flash drives,
FAT 32
LXI compliance
LXI version 1.4 Core 2011
Language
Embedded Test Script Processor (TSP) accessible from any host
interface; responds to high-speed test scripts comprised of remote
commands and statements (for example, branching, looping, math);
able to execute high-speed test scripts stored in memory without
host intervention; also SCPI (default command set)
Accessories supplied
Product Information CD-ROM, Model DMM7510 Quick Start Guide,
Kickstart Software Quick Start Guide, power cord, 1 m USB cable
(type A to type B), 3 m LAN cable, and Model 1756 Standard Test
Lead Kit
Accessories available
(Calibration / Data / ISO 17025), software IVI/VISA drivers for
Microsoft
®
Visual Basic
®
, Visual C/C++
®
, National Instruments
(NI™) LabVIEW, Keithley Test Script Builder, Keithley KickStart,
and NI LabWindows/CVI
Display
Five-inch capacitive touch, color thin-film-transistor (TFT) WVGA
(800 x 480) with LED backlight
Password protection
30 characters
Expansion interface
The TSP-Link
®
expansion interface allows TSP-enabled
instruments to trigger and communicate with each other
IP configuration
Static or DHCP (manual or automatic)