Datasheet

Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications and characteristics are subject to change without notice.
Page 22 of 25 SPEC-DMM7510 Rev. B / October 2016
SYSTEM PERFORMANCE, TYPICAL
Mode: 3½ digit, autozero off, 0.0005 PLC, excludes measurement time
Time includes function change from DC voltage or 2-wire ohms to listed function
RANGES FOR FUNCTION CHANGE TIMES
Function change times apply to the ranges listed in the table below.
Buffer transfer speed (binary) Measurements into computer (per second)
USB LAN GPIB
Average for 1000 readings 280000 270000 190000
Average for 1000 readings with timestamp 170000 140000 100000
95
For DC voltage or 2-wire ohms to frequency or period, 10 ms aperture. For AC current or AC voltage, detector bandwidth is 300 Hz.
Function Function change (ms) Range change (ms)
DC voltage or 2-wire ohms (< 10 k)
6 1.3
4-wire ohms (< 10 k)
7 1.3
DC current 7 1.3
Frequency or period
95
7 1.3
AC voltage or AC current 7 1.3
Digitize voltage or current 7 1.3
Function Range
DC voltage 10 V
2-wire or 4-wire ohms
1 k
DC current 1 mA
Dry-circuit ohms
10
Thermocouple Use DC voltage rates
Thermistor Use 2-wire ohms rates
AC current 1 mA
AC voltage 1 V