Datasheet

Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications and characteristics are subject to change without notice.
Page 2 of 25 SPEC-DMM7510 Rev. B / October 2016
RMS NOISE (ADDITIONAL PEAK NOISE UNCERTAINTY)
7
Applies to ± ppm of range
Peak noise uncertainty is included in DC specifications for ≥ 1 PLC
Add peak noise uncertainty to measurements for < 1 PLC
Input impedance set to auto
Examples:
10 V at 0.006 PLC: 1.2 (from Accuracy table) + 11 (additional peak noise uncertainty) = 12.2 ppm of range
10 V at 1 PLC: 1.2 + 0 = 1.2 ppm of range
NPLC Digits 100 mV 1 V 10 V 100 V 1000 V
5 0.5 0.08 0.06 0.3 0.06
1 0.5 0.09 0.07 0.4 0.07
0.2
8
2 (10) 0.2 (1.6) 0.1 (1.1) 1.1 (9.4) 0.1 (1)
0.2 2 (12) 0.2 (1.6) 0.1 (1) 1.1 (8.9) 0.2 (1.1)
0.06 3 (17) 0.4 (2.7) 0.3 (2.1) 3 (17) 0.3 (2.4)
0.006 6 (42) 3 (18) 1 (11) 20 (100) 3 (18)
0.0005 30 (220) 20 (150) 20 (130) 120 (690) 20 (150)
DC VOLTAGE SENSE ACCURACY
Range Accuracy ±[ppm of reading + ppm of range]
24 hour
T
CAL
±1 °C
90 day
T
CAL
±5 °C
1 year
T
CAL
±5 °C
2 year
T
CAL
±5 °C
Temperature
coefficient
9
100.00000 mV 6 + 14 12 + 14 18 + 14 29 + 14 0.1 + 2.5
1.0000000 V 4 + 1.5 9 + 3 15 + 3 26 + 3 0.1 + 0.5
10.000000 V 2 + 1.0 9 + 1.8 14 + 1.8 22 + 1.8 0.1 + 0.05
DC VOLTAGE RATIO
For input signals ≥ 1 % of the range,
ratio accuracy =
±[[V
INPUT
ppm of reading + V
INPUT
ppm of range * (V
INPUT
range/V
INPUT
input)]
+ [V
SENSE
ppm of reading + V
SENSE
ppm of range * (V
SENSE
range/V
SENSE
input)]]
DC VOLTAGE CHARACTERISTICS
ADC linearity
1.0 ppm of reading + 1.0 ppm of range
Input impedance
100 mV to 10 V ranges: Selectable > 10 G ǁ < 400 pF (auto) or
10 M ±1 % (10 M)
100 V to 1000 V ranges: 10 M ±1 %
Input bias current
< 50 pA at 23 °C under the following conditions: Autozero off or input
impedance 10 M
Common mode current
< 2.1 µA peak-peak in 1 MHz bandwidth
< 100 nA peak-peak in 1 kHz bandwidth
Common mode voltage
500 V
peak
LO terminal to chassis maximum
DC voltage autozero off error
For ±1 °C and ≤ 10 minutes, add ± (8 ppm of reading + 15 µV)
7
Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. V
RMS
noise is typical. Additional peak noise is
guaranteed.
8
With line sync on.
9
Add per degree from T
CAL
± 5
o