Datasheet

Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications and characteristics are subject to change without notice.
Page 18 of 25 SPEC-DMM7510 Rev. B / October 2016
TRUE RMS AC VOLTAGE AND AC CURRENT
Function Range
83
Resolution
1 year accuracy: ± (% of reading + % of range) T
CAL
± 5 °C
3 Hz to 5 Hz
5 Hz to
10 Hz
10 Hz to
20 kHz
20 kHz to
50 kHz
50 kHz to
100 kHz
100 kHz to
300 kHz
Voltage
84
100.0000 mV
0.1 µV
1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5
1.000000 V
1 µV
1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5
10.00000 V
10 µV
1.0 + 0.03 0.30 + 0.03 0.06 + 0.03
0.14 + 0.05
0.6 + 0.08 4.0 + 0.5
100.0000 V
100 µV
1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5
700.000 V 1 mV 1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5
Temperature
coefficient /
°C (all
ranges)
0.01 + 0.003 0.03 + 0.003 0.005 + 0.003 0.006 + 0.005 0.01 + 0.006 0.03 + 0.01
Function
Range
83
Resolution 1 year accuracy: ± (% of reading + % of range) T
CAL
± 5 °C
3 Hz to 5 Hz 5 Hz to 10 Hz 10 Hz to 2 kHz 2 kHz to 5 kHz 5 kHz to 10 kHz
Current
84
1.000000 mA 1 nA 1.0 + 0.04 0.30 + 0.04 0.08 + 0.03 0.09 + 0.03 0.09 + 0.03
10.00000 mA 10 nA 1.0 + 0.04 0.30 + 0.04 0.08 + 0.03 0.09 + 0.03 0.09 + 0.03
100.0000 mA 100 nA 1.0 + 0.04 0.30 + 0.04 0.08 + 0.03 0.09 + 0.03 0.09 + 0.03
1.000000 A
1 µA
1.0 + 0.04 0.30 + 0.04 0.20 + 0.04 0.88 + 0.04 2.0 + 0.04
3.000000 A
1 µA
1.0 + 0.05 0.30 + 0.05 0.20 + 0.05 0. 88 + 0.05 2.0 + 0.05
10.00000 A
85
10 µA 1.0 + 0.05 0.40 + 0.05 0.40 + 0.05 0. 88 + 0.05 2.0 + 0.05
Temperature
coefficient /
°C (all
ranges)
0.10 + 0.004 0.030 + 0.004 0.005 + 0.003 0.006 + 0.005 0.006 + 0.005
ADDITIONAL AC UNCERTAINTIES – LOW FREQUENCY UNCERTAINTY
Additional uncertainty ±(% of reading),
lower frequency uncertainty
Detector bandwidth (BW)
3 BW (3 Hz to 300 kHz) 30 BW (30 Hz to 300 kHz) 300 BW (300 Hz to 300 kHz)
20 Hz to 30 Hz 0 0.3
30 Hz to 50 Hz 0 0
50 Hz to 100 Hz 0 0 4.0
100 Hz to 200 Hz 0 0 0.72
200 Hz to 300 Hz 0 0 0.18
300 Hz to 500 Hz 0 0 0.07
> 500 Hz 0 0 0
83
20 % overrange on AC functions except 1 % on 700 V, 3.33 % on 3 A and 1 % on 10 A. Default resolution is 6½ digits.
84
Specifications are for detector bandwidth of 3 Hz and sine wave inputs > 5 % of range. Detector bandwidth of 3 Hz and 30 Hz are multisample
A/D conversions. Detector bandwidth of 300 Hz is a single A/D conversion, programmable from 0.0005 PLC to 15 PLC (60 Hz), 12 PLC
(50 Hz). Default condition set to 1 PLC.
85
Rear input terminals only.