Datasheet
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications are subject to change without notice.
SPEC-DMM7510 Rev. B / October 2016 Page 13 of 25
TYPICAL AC COUPLED
Range
Analog
bandwidth
(-3 dB)
Maximum flatness
error
3 Hz to 20 kHz
64
Filter Fast
settling time
(0.5 %)
Filter Slow
settling time
(0.5 %)
Low frequency
coupling point
66
(-3 dB)
100.000 V 600 kHz 0.1 dB 80 ms 2.3 s 1 Hz
1000.00 V
600 kHz 0.1 dB 80 ms 2.3 s
1 Hz
DC-COUPLED ADDITIONAL NOISE UNCERTAINTY, TYPICAL
68
68
Specified with aperture Auto and 4-wire short on input terminals. For 100 V range, input impedance 10 MΩ, multiply by 2.5. For all ranges and
sample rate > 1 k, add an additional 3× RMS noise uncertainty to ppm of range.
0.1
1
10
100
1000
1k 2k 5k 10k 20k 50k 100k 200k 500k 1M
ppm of range
Samples per second
RMS noise, digitize voltage
100mV
1V
10V
100V
1000V