Datasheet
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications
Specifications are subject to change without notice.
SPEC-DMM7510 Rev. B / October 2016 Page 11 of 25
CONTINUITY
Range
49
Resolution Test
current
Open circuit
voltage
Accuracy ±[ppm of reading + ppm of range]
2 year
T
CAL
±5 °C
Temperature coefficient
50
1.0000 kΩ 100 mΩ
1 mA 14.0 V 100 + 100 2.5 + 1
CONTINUITY CHARACTERISTICS
Continuity high limit
User-selectable; default 10 Ω
CAPACITANCE
Accuracies specified for additional cable and stray capacitance properly zeroed with the Rel function.
ACCURACY
Range
51
Resolution Charge
current
52, 53
Maximum
circuit
voltage
Accuracy ±[% of reading + % of range]
2 year
T
CAL
±5 °C
Temperature coefficient
50
1.0000 nF 0.1 pF 1.1 µA 2.8 V 1 + 0.2 0.15 + 0.05
10.000 nF 1 pF 1.1 µA 2.8 V 1 + 0.1 0.15 + 0.01
100.00 nF 10 pF 10 µA 3 V 0.4 + 0.1 0.01 + 0.01
1.0000 µF 0.1 nF 100 µA 3 V 0.4 + 0.1 0.01 + 0.01
10.000 µF 1 nF 100 µA 3 V 0.4 + 0.1 0.01 + 0.01
100.00 µF 10 nF 1 mA 3 V 0.4 + 0.1 0.01 + 0.01
1000.0 µF 0.1 µF 10 mA 3 V 0.5 + 0.1 0.01 + 0.01
DIODE
Voltage
measure
range
51
Resolution Bias level
(selectable)
Accuracy ±[ppm of reading + ppm of range]
90 day
T
CAL
±5 °C
1 year
T
CAL
±5 °C
2 year
T
CAL
±5 °C
Temperature
coefficient
50
10.000000 V 1 µV 10 µA / 100 µA / 1 mA 20 + 5 30 + 5 45 + 5 2.5 + 1
49
Specifications exclude lead resistance.
50
Add per degree from T
CAL
±5
o
C; specifications without autocalibration.
51
20 % overrange on all ranges.
52
Charging current values are typical, guaranteed by design.
53
Discharge current limited to < 13 mA.