7½-Digit Graphical Sampling Multimeter
RMS NOISE (additional peak noise uncertainty)
36
Applies to ± ppm of range.
Peak noise uncertainty is included in DC specifications for 1 PLC.
Add peak noise uncertainty to measurements when < 1 PLC.
EXAMPLES:
10 at 0.2 PLC: 80 (from Accuracy table) + 230 (additional peak noise uncertainty) = 310 ppm
of range.
10 at 1 PLC: 80 + 0 = 80 ppm of range.
NPLC Digits 1 10 100 1 k 10 k
5 10 11 6 5 0.9
1 9 9 7 7 0.8
0.2
37
30 (130) 30 (120) 30 (120) 30 (120) 3 (16)
0.2 60 (220) 60 (230) 50 (190) 50 (190) 9 (35)
0.06 70 (350) 70 (350) 50 (290) 50 (280) 20 (90)
0.006 130 (750) 120 (830) 110 (700) 100 (690) 20 (110)
0.0005 520 (3550) 530 (3520) 530 (3380) 500 (3370) 100 (670)
DRY CIRCUIT RESISTANCE CHARACTERISTICS
MAXIMUM 4-WIRE OHMS LEAD RESISTANCE:
0.5 per lead for 1 range.
10% of range per lead for 10 to 100 ranges.
50 per lead for 1k to 10k ranges.
INPUT CURRENT LIMIT: For signals greater than ±20mV, current limited, ±13mA typical.
OFFSET COMPENSATION: Selectable on 1 to 10k ranges.
AUTOZERO OFF ERROR: For ±1°C and 10 minutes, add ±8 ppm of reading.
DC Current
ENHANCED ACCURACY (within 30 days of autocalibration, T
OPER
±5°C from T
ACAL
)
Range
38
Resolution
Maximum
Burden Voltage
Accuracy ±(ppm of reading + ppm of range)
24 Hour
T
CAL
±1°C
39
90 Day
T
CAL
±5°C
1 Year
T
CAL
±5°C
2 Year
T
CAL
±5°C
Temperature
Coefficient
40
10.000000 µA 1 pA 15 mV 30 + 30 75 + 30 75 + 30 75 + 30 0.15 + 0.1
100.00000 µA 10 pA 15 mV 20 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0.1
1.0000000 mA 100 pA 15 mV 30 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0.1
10.000000 mA 1 nA 20 mV 40 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0.1
100.00000 mA 10 nA 200 mV 50 + 18 150 + 30 150 + 30 150 + 30 0.15 + 0.1
1.0000000 A 100 nA 400 mV 150 + 50 400 + 50 400 + 50 400 + 50 0.15 + 0.1
3.000000 A 1 µA 1300 mV 200 + 40 400 + 40 400 + 40 400 + 40 0.15 + 0.1
10.000000 A
41
1 µA 650 mV 700 + 275 800 + 275 1500 + 275 2000 + 275 50 + 10
ACCURACY
Range
38
Resolution
Maximum
Burden
Voltage
Accuracy ±(ppm of reading + ppm of range)
24 Hour
T
CAL
±1°C
39
90 Day
T
CAL
±5°C
1 Year
T
CAL
±5°C
2 Year
T
CAL
±5°C
Temperature
Coefficient
40
10.000000 µA 1 pA 15 mV 30 + 30 100 + 30 125 + 40 175 + 50 10 + 8
100.00000 µA 10 pA 15 mV 20 + 5 75 + 12 100 + 15 150 + 20 10 + 3
1.0000000 mA 100 pA 15 mV 30 + 5 75 + 12 100 + 15 150 + 20 10 + 3
10.000000 mA 1 nA 20 mV 40 + 5 75 + 12 100 + 15 150 + 20 10 + 3
100.00000 mA 10 nA 200 mV 50 + 18 300 + 30 400 + 30 500 + 30 50 + 5
1.0000000 A 100 nA 400 mV 150 + 50 400 + 50 450 + 50 500 + 50 10 + 10
3.000000 A 1 µA 1300 mV 200 + 40 400 + 40 450 + 40 500 + 40 10 + 10
10.000000 A
41
1 µA 650 mV 700 + 275 800 + 275 1500 + 275 2000 + 275 50 + 10
36. Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. RMS noise is typical. Additional peak noise is guaranteed.
37. With line sync on.
38. 20% overrange supported for all ranges except for 3A and 10A, which are 1% supported.
39. Relative to calibration accuracy.
40. Add per degree from T
CAL
±5°C.
41. Rear input terminals only.
Model DMM7510 condensed specifications