Datasheet

DIGITAL MULTIMETERS & SYSTEMS
www.keithley.com
1.888.KEITHLEY
(U.S. only)
A Greater Measure of Confidence
A Tektronix Company
DMM7510
7½-Digit Graphical Sampling Multimeter
RESISTANCE OPEN CIRCUIT DC VOLTAGE
28
Range
20
2-wire
Offset compensation off Offset compensation on
4-wire 4-wire
1 9.2 V 9.5 V
10 9.2 V 9.2 V 9.5 V
100 , 1 k 14.0 V 14.2 V 14.3 V
10 k 9.5 V 9.5 V 0.0 V
100 k, 1 M 12.7 V 14.3 V 0.0 V (100 k range only)
10 M to 1 G 6.9 V 6.9 V
4-WIRE OHMS (10k) Offset Compensation On
RMS NOISE (additional peak noise uncertainty)
29
Applies to ± ppm of range.
Peak noise uncertainty is included in DC specifications for 1 PLC.
Add peak noise uncertainty to measurements for <1 PLC.
EXAMPLES
1k at 0.006 PLC: 3 (from Accuracy table) + 26 (additional peak noise uncertainty) = 29 ppm
of range.
1k at 1 PLC: 3 + 0 = 3 ppm of range.
NPLC Digits 1 10 100 1 k 10 k
5 2.8 0.3 0.3 0.07 0.3
1 4.2 0.4 0.4 0.12 0.5
0.2
30
30 (160) 3 (13) 3 (13) 0.4 (2.6) 1.2 (8.2)
0.2 50 (250) 5 (22) 5 (22) 0.6 (3.2) 1.2 (8.3)
0.06 110 (490) 11 (47) 11 (46) 1.1 (6.6) 2 (16)
0.006 110 (710) 10 (70) 10 (70) 4 (26) 10 (60)
0.0005 520 (3420) 50 (340) 50 (340) 40 (220) 50 (300)
2-WIRE OHMS
RMS NOISE (additional peak noise uncertainty)
29
Applies to ± ppm of range.
Peak noise uncertainty is included in DC specifications for 1 PLC.
Add peak noise uncertainty to measurements for <1 PLC.
EXAMPLES
10k at 0.006 PLC: 3 (from Accuracy table) + 5 (50m with Rel ) + 43 (additional peak noise
uncertainty) = 51 ppm of range.
10k at 1 PLC: 3 + 5 + 0 = 8 ppm of range.
NPLC Digits 10 100 1 k 10 k
5 1.1 0.8 0.1 0.2
1 0.6 0.6 0.09 0.4
0.2
30
2 (17) 2 (10) 0.2 (1.5) 0.8 (6.3)
0.2 2 (17) 2 (14) 0.3 (1.6) 0.8 (6.4)
0.06 3 (22) 3 (19) 0.4 (3.7) 2 (12)
0.006 6 (50) 6 (50) 3 (21) 6 (43)
0.0005 30 (300) 30 (230) 20 (150) 30 (210)
RESISTANCE CHARACTERISTICS
MAXIMUM 4-WIRE OHMS LEAD RESISTANCE: 5 per lead for 1 range, 10% of range per lead
for 10 to 1k ranges; 1k per lead for all other ranges.
OFFSET COMPENSATION: Selectable on 4-wire, 1 to 100k ranges.
OPEN LEAD DETECTOR: Default is off.
AUTOZERO OFF ERROR:
For 2-wire ohms, ±1°C and 10 minutes, add ±(8ppm of reading) + 1.5m for 10, 15m
for 100 and 1k ranges, 150m for 10k range, 1.5 for 100 k range, and 15 for all
other ranges.
For 4-wire ohms, ±1°C and 10 minutes, add ±(8ppm of reading).
INPUT CURRENT LIMIT:
For signals with a magnitude of +12V to +40V or –12V to –40V: ±13mA source or sink, typical.
For signals with a magnitude of greater than +40V or –40V: ±130µA source or sink, typical.
Dry Circuit Resistance
ENHANCED ACCURACY (within 30 days of autocalibration, T
OPER
±5ºC from T
ACAL
)
Range
31
Resolution
Test Current
35
(±5%)
Open Circuit
DUT Voltage
32
Accuracy ±(ppm of reading + ppm of range)
24 Hour
T
CAL
±1°C
33
90 Day
T
CAL
±5°C
1 Year
T
CAL
±5°C
2 years
T
CAL
±5°C
Temperature
Coefficient
34
1.000000 1 µ 10 mA 25 mV 25 + 80 50 + 80 50 + 80 50 + 80 1.5 + 0.1
10.00000 10 µ 1 mA 25 mV 25 + 80 50 + 80 50 + 80 50 + 80 1.5 + 0.1
100.0000 100 µ 100 µA 25 mV 25 + 80 90 + 80 90 + 80 90 + 80 1.5 + 0.1
1.000000 k 1 m 10 µA 25 mV 25 + 80 180 + 80 180 + 80 180 + 80 1.5 + 0.1
10.00000 k 10 m 5 µA 25 mV 25 + 80 320 + 80 320 + 80 320 + 80 1.5 + 0.1
ACCURACY
Range
31
Resolution
Test Current
35
(±5%)
Open Circuit
DUT Voltage
32
Accuracy ±(ppm of reading + ppm of range)
24 Hour
T
CAL
±1°C
33
90 Day
T
CAL
±°C
1 Year
T
CAL
±5°C
2 Year
T
CAL
±5°C
Temperature
Coefficient
34
1.000000 1 µ 10 mA 25 mV 25 + 80 50 + 80 70 + 80 90 + 80 2.5 + 1
10.00000 10 µ 1 mA 25 mV 25 + 80 50 + 80 70 + 80 90 + 80 5 + 1
100.0000 100 µ 100 µA 25 mV 25 + 80 90 + 80 140 + 80 200 + 80 2.5 + 1
1.000000 k 1 m 10 µA 25 mV 25 + 80 180 + 80 400 + 80 600 + 80 5 + 1
10.00000 k 10 m 5 µA 25 mV 25 + 80 320 + 80 800 + 80 1300 + 80 8 + 1
28. Open circuit voltage is typical, measured from input HI to LO, SHI and SLO open. For 1 to
1M ranges using an external digital multimeter (DMM) set to 10M input impedance; for
10M to 1G ranges, set external DMM to >10G input impedance.
29. Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short.
RMS noise is typical. Additional peak noise is guaranteed.
30. With line sync on.
31. 20% overrange on all ranges, except 2.4k for the 10k range.
32. Maximum clamp voltages are DC, typical accuracy is ±20%. Add 20% for offset compensation on.
33. Relative to calibration accuracy.
34. Add per degree from T
CAL
±5°C.
35. Test current with offset compensation off.
Model DMM7510 condensed specifications
Model DMM7510 condensed specifications