Datasheet
DIGITAL MULTIMETERS & SYSTEMS
A Greater Measure of Confidence
www.keithley.com
1.888.KEITHLEY
(U.S. only)
A Tektronix Company
DMM7510
7½-Digit Graphical Sampling Multimeter
DC VOLTAGE CHARACTERISTICS
ADC LINEARITY: 1.0 ppm of reading + 1.0 ppm of range.
INPUT IMPEDANCE:
100mV to 10V Ranges: Selectable >10GΩ
||
<400pF (auto) or 10MΩ ±1% (10MΩ).
100V to 1000V Ranges: 10MΩ ±1%.
INPUT BIAS CURRENT: <50pA at 23°C under the following conditions: Autozero off or input
impedance 10MΩ.
COMMON MODE CURRENT: <2.1µA peak-peak in 1MHz bandwidth.
<100nA peak-peak in 1kHz bandwidth.
COMMON MODE VOLTAGE: 500V
peak
LO terminal to chassis maximum.
DC VOLTAGE AUTOZERO OFF ERROR:
For ±1°C and ≤10 minutes, add ±(8ppm of reading + 15µV).
10. Specifications are for 4-wire resistance, offset compensation on for ≤10kΩ measurements, and
offset compensation off for ≥10kΩ measurements. 1Ω range is 4-wire only. For 2-wire, with
Rel, add 50mΩ to ppm of range uncertainty. Without Rel and with Model 1756 test leads, add
100mΩ to ppm of range uncertainty.
11. 20% overrange on all ranges.
12. Test current with offset compensation off, ±5%.
13. Relative to calibration accuracy.
14. Add per degree from T
CAL
±5°C.
15. Specifications are for external cable and load capacitance <1nF.
16. For offset compensation on, add 10ppm uncertainty to ppm of reading.
17. For 4-wire 1MΩ, open lead detector on, add 10 ppm uncertainty to ppm of reading.
18. Specified for <10% lead resistance mismatch in HI and LO.
19. Specifications are for 4-wire resistance, offset compensation on for ≤10kΩ measurements, and
offset compensation off for ≥10kΩ measurements. 1Ω range is 4-wire only. For 2-wire, with
Rel, add 50mΩ to ppm of range uncertainty. Without Rel and with Model 1756 test leads, add
100mΩ to ppm of range uncertainty.
20. 20% overrange on all ranges.
21. Test current with offset compensation off.
22. Relative to calibration accuracy.
23. Add per degree from T
CAL
±5°C.
24. Specifications are for external cable and load capacitance <1nF.
25. For offset compensation on, add 10ppm of uncertainty to ppm of reading.
26. For 4-wire, 1MΩ, open lead detection on, add 10ppm uncertainty to ppm of reading.
27. Specified for <10% lead resistance mismatch in HI and LO.
Resistance
ENHANCED ACCURACY (within 30 days of autocalibration, T
OPER
±5°C from T
ACAL
)
10
Range
11
Resolution
Test Current
12
(±5%)
Accuracy ±(ppm of reading + ppm of range)
24 Hour
T
CAL
±1°C
13
90 Day
T
CAL
±5°C
1 Year
T
CAL
±5°C
2 Year
T
CAL
±5°C
Temperature
Coefficient
14
1.0000000 Ω 0.1 µΩ 10 mA 15 + 50 30 + 50 30 + 50 30 + 50 0.15 + 0.1
10.000000 Ω 1 µΩ 10 mA 15 + 5 30 + 5 30 + 5 30 + 5 0.15 + 0.1
100.00000 Ω 10 µΩ 1 mA 12 + 4 27 + 4 27 + 4 27 + 4 0.15 + 0.1
1.0000000 kΩ 100 µΩ 1 mA 12 + 3 24 + 3 24 + 3 24 + 3 0.15 + 0.1
10.000000 kΩ
15
1 mΩ 100 µA 13 + 3 30 + 3 30 + 3 30 + 3 0.15 + 0.1
100.00000 kΩ
15, 16
10 mΩ 10 µA 13 + 3 30 + 3 30 + 3 30 + 3 0.15 + 0.1
1.0000000 MΩ
15, 17
100 mΩ 10 µA 14 + 3 30 + 4 30 + 4 30 + 4 0.15 + 0.1
10.000000 MΩ
18
1 Ω 0.69 µA
||
10 MΩ 150 + 6 200 + 10 200 + 10 200 + 10 70 + 1
100.00000 MΩ
18
10 Ω 0.69 µA
||
10 MΩ 800 + 30 2000 + 30 2000 + 30 2000 + 30 385 + 1
1.0000000 GΩ
18
100 Ω 0.69 µA
||
10 MΩ 9000 + 100 9000 + 100 9000 + 100 9000 + 100 3000 + 1
ACCURACY
19
Range
20
Resolution
Test Current
21
(±5%)
Accuracy ±(ppm of reading + ppm of range)
24 Hour
T
CAL
±1°C
22
90 Day
T
CAL
±5°C
1 Year
T
CAL
±5°C
2 Year
T
CAL
±5°C
Temperature
Coefficient
23
1 Ω 0.1 µΩ 10 mA 15 + 50 40 + 50 50 + 50 70 + 50 2.5 + 5
10 Ω 1 µΩ 1 mA 15 + 5 40 + 5 50 + 5 70 + 5 2.5 + 0.5
100 Ω 10 µΩ 1 mA 12 + 4 35 + 4 47 + 4 65 + 4 5 + 0.25
1 kΩ 100 µΩ 1 mA 12 + 3 30 + 3 41 + 3 65 + 3 5 + 0.25
10 kΩ
24
1 mΩ 100 µA 10 + 3 30 + 3 42 + 3 65 + 3 2.5 + 0.25
100 kΩ
24, 25
10 mΩ 10 µA 13 + 3 38 + 3 50 + 3 65 + 3 5 + 1
1 MΩ
24, 26
100 mΩ 10 µA 14 + 3 38 + 5 50 + 5 65 + 5 5 + 1
10 MΩ
27
1 Ω 0.69 µA
||
10 MΩ 150 + 6 200 + 10 400 + 10 600 + 12 70 + 1
100 MΩ
27
10 Ω 0.69 µA
||
10 MΩ 800 + 30 2000 + 30 2000 + 30 2600 + 30 385 + 1
1 GΩ
27
100 Ω 0.69 µA
||
10 MΩ 9000 + 200 9000 + 200 13000 + 200 14000 + 200 3000 + 1
NORMAL MODE REJECTION
For DC voltage, line frequency ±0.1%.
5 PLC 1 PLC ≤0.2 PLC ≤0.01 PLC
Line Sync On
110 dB 90 dB 45 dB —
Line Sync Off
60 dB 60 dB — —
COMMON MODE REJECTION
For DC voltage and 1kΩ unbalanced in LO terminal; AC CMRR is 70dB.
NPLC 5 1 0.2 ≤ 0.2
Line Sync
On On On Off
CMRR
140 dB 140 dB 120 dB 80 dB
Model DMM7510 condensed specifications
Model DMM7510 condensed specifications