87. Rear input terminals only.
88. Specified for square wave inputs. Input signal must be >10% of ACV range. If input is<20mV
on the 100mV range, then the frequency must be >10Hz. For sine wave inputs, frequency
must be >100Hz. For frequencies 100Hz, threshold level 50% of input signal and 7Hz,
threshold level 3% of range.
89. Threshold range is voltage RMS and threshold level voltage peak. Specified with 1kHz square
wave. 100V and 700V threshold ranges guaranteed by design.
90. Reading speeds for autozero off, fixed range, autodelay off. Offset compensation off and open
lead detector off where applicable.
91. Buffer measurements: For <0.2 PLC, multisample, single buffer transfer binary reading only.
92. PC measurements: For 1 and 0.2 PLC single reading and single transfer to computer (USB).
93. Reading rates using factory default operating conditions and autorange off, autodelay off.
Speeds include measurement and data transfer out of the USB. 1000 readings with binary
transfer over USB.
94. For bandwidth 300Hz, autozero off, 6½ digits at 1 PLC, 3½ digits at 0.0005 PLC.
AC VOLTAGE CHARACTERISTICS
MEASUREMENT METHOD: AC-coupled, true RMS.
INPUT IMPEDANCE: 1M ± 2%
||
<150pF.
VOLT*HERTZ PRODUCT: <2.1 × 10
7
V*Hz verified; input frequency verified for <300kHz.
DMM7510
7½-Digit Graphical Sampling Multimeter
AC CURRENT CHARACTERISTICS
MEASUREMENT METHOD: AC-coupled, true RMS.
Range 1 mA 10 mA 100 mA 1 A 3 A 10 A
87
Burden Voltage (RMS)
<16 mV <20 mV <0.2 V <0.4 V <1.3 V <0.65 V
Overload Recovery: For
each additional sustained
ampere beyond ±1.5A, add
the following initial % of
range error until thermally
settled after overload
recovery
0.006 0.006 0.12 0.05
Frequency and Period
MEASUREMENT ACCURACY
88
Aperture
Measurement
Resolution
Accuracy
±(ppm of reading + ppm of aperture time)
Frequency: 3 Hz to 500 kHz
Period: 333 ms to 2 µs
1 Year, T
CAL
±5°C 2 Year, T
CAL
±5°C
250 ms 0.1 ppm 80 + 0.333 160 + 0.333
100 ms 0.1 ppm 80 + 3.33 160 + 3.33
10 ms 0.1 ppm 80 + 33.3 160 + 33.3
THRESHOLD LEVEL ACCURACY
89
Threshold Range
Threshold
Resolution
Accuracy ±(% of reading)
2 Year, T
CAL
±5°C
100 mV to 700 V 0.05% 1.0%
FREQUENCY AND PERIOD CHARACTERISTICS
MEASUREMENT METHOD: Reciprocal counting technique.
APERTURE: 10ms to 273ms; default is 10ms.
TYPICAL READING RATES, 60Hz (50Hz) OPERATION
90, 91, 92, 93
Functions:
DC Voltage (10 V),
2-wire Ohms (10k),
DC Current (1 mA)
Functions:
4-wire ohms (1 k),
4-wire/3-wire RTD
Functions:
Thermistor
Functions:
Dry Circuit (1 k)
NPLC Digits
Measurements
Into Buffer
Measurements
Into Computer
Measurements
Into Buffer
Measurements
Into Computer
Measurements
Into Buffer
Measurements
Into Computer
Measurements
Into Buffer
Measurements
Into Computer
1 59.8 (49.8) 58 (48) 29 (24) 28 (24) 57 (48) 57 (48) 27 (23) 26 (22)
0.2 295 (240) 250 (210) 128 (109) 119 (100) 230 (200) 230 (200) 100 (89) 96 (85)
0.06 965 (810) 950 (800) 310 (280) 315 (280) 900 (750) 900 (750) 190 (180) 190 (180)
0.006 7500 (6700) 7300 (6500) 750 (730) 740 (720) 6800 (6000) 6800 (6000) 295 (290) 295 (290)
0.0005 26000 (26000) 24000 (24000) 860 (860) 860 (860) 18000 (18000) 18000 (18000) 310 (310) 310 (310)
Functions: ACV, ACI
Detector Bandwidth
(Hz) Digits
Measurements
Into Buffer
Measurements
Into Computer
3 0.5 (0.5) 0.5 (0.5)
30 3.3 (3.3) 3.3 (3.3)
300
94
59.8 (49.8) 55 (46)
300
94
26200 (26200) 24500 (24500)
DIGITIZE, TYPICAL
Sampling Rate Digits Resolution
Measurements Into
Computer
93
10 kS/s 18 9700
20 kS/s 16 19000
50 kS/s 16 44400
100 kS/s 15 80000
1 MS/s 12 108000
Model DMM7510 condensed specifications
Model DMM7510 condensed specifications