Datasheet

DIGITAL MULTIMETERS & SYSTEMS
www.keithley.com
1.888.KEITHLEY
(U.S. only)
A Greater Measure of Confidence
A Tektronix Company
DMM7510
7½-Digit Graphical Sampling Multimeter
RMS NOISE (additional peak noise uncertainty)
42
Applies to ± ppm of range.
Peak noise uncertainty is included in DC specifications for 1 PLC.
Add peak noise uncertainty to measurements when <1 PLC.
EXAMPLES:
1mA at 0.006 PLC: 9 (from Accuracy table) + 20 (additional peak noise uncertainty) = 29 ppm of range.
1mA at 1 PLC: 9 +0 = 9 ppm of range.
NPLC Digits 10 µA 100 µA 1 mA 10 mA 100 mA 1A 3A 10A
43
5 0.15 0.14 0.09 0.1 0.3 0.3 0.2 0.8
1 0.4 0.13 0.1 0.1 0.5 0.5 0.3 1.2
0.2 0 (220) 0 (23) 0.2 (3.4) 0.2 (1.6) 2 (10) 2 (11) 0.7 (4.6) 4 (32)
0.2
44
120 (260) 12 (26) 1.2 (3.8) 0.3 (1.8) 1.9 (9.8) 2 (10) 0.8 (5) 8 (37)
0.06 130 (280) 12 (29) 1.3 (5.6) 0.4 (3.9) 2 (14) 2 (14) 1.2 (7.7) 10 (59)
0.006 130 (350) 14 (42) 3 (20) 2 (20) 4 (30) 4 (31) 7 (51) 20 (110)
0.0005 260 (2110) 30 (300) 20 (150) 20 (160) 30 (190) 30 (190) 70 (510) 60 (420)
DC CURRENT CHARACTERISTICS
Range 10 µA 100 µA 1 mA 10 mA 100 mA 1 A 3 A 10 A
43
Effective Internal Shunt Value
45
1 k 100 10 1 0.1 0.1 0.1 0.005
Autozero Off Error: For ±1°C and 10 minutes add
±(8 ppm of reading + range error)
150 pA 1.5 nA 15 nA 150 nA 15 µA 150 µA 150 µA 3 mA
Overload Recovery: For each additional sustained amp
beyond ±1.5A, add the following initial ppm of range error
until thermally settled after overload recovery
15500 1800 150 150 6500 200
Temperature
4-WIRE RTD OR 3-WIRE RTD
TYPES: 100 platinum PT100, D100, F100, PT385, PT3916; or user-configurable 0 to 10k.
Type Range Resolution
Accuracy ±°C
2 Year
T
CAL
±5°C
Temperature
Coefficient
46
4-Wire RTD
–200 to 850 °C 0.01 °C 0.06 °C 0.003 °C/°C
3-Wire RTD
47
–200 to 850 °C 0.01 °C 0.75 °C 0.003 °C/°C
THERMISTOR
TYPES: 2.252k, 5k, and 10k.
Type Range Resolution
Accuracy ±°C
2 Year
T
CAL
±5°C
Temperature
Coefficient
46
Thermistor
80 to +150 °C 0.01 °C 0.08 °C 0.002 °C/°C
THERMOCOUPLE
TYPES: B, E, J, K, N, R, S, T
Accuracy ±°C
Type Range Resolution
2 Year, T
CAL
±5°C
48
Simulated Reference
Junction
Temperature
Coefficient
46
B 350 to +1820 °C 0.1 °C 0.6 °C 0.03 °C/°C
E −200 to +1000 °C 0.001 °C 0.2 °C 0.03 °C/°C
J 200 to +760 °C 0.001 °C 0.2 °C 0.03 °C/°C
K 200 to +1372 °C 0.001 °C 0.2 °C 0.03 °C/°C
N −200 to +1300 °C 0.001 °C 0.2 °C 0.03 °C/°C
R 0 to +1768 °C 0.1 °C 0.6 °C 0.03 °C/°C
S 0 to +1768 °C 0.1 °C 0.6 °C 0.03 °C/°C
T 100 to +400 °C 0.001 °C 0.2 °C 0.03 °C/°C
42. Noise values are based on 1000 readings with autozero on and AMPS terminal open. RMS
noise is typical. Additional peak noise is guaranteed.
43. Rear input terminals only.
44. With line sync on.
45. Values are typical and guaranteed by design.
46. Add per degree from T
CAL
±5°C; specifications without autocalibration.
47. For 3-wire RTD, accuracy is for <0.1 lead resistance mismatch for input HI and LO. Add
0.25°C/0.1 of HI-LO lead resistance mismatch.
48. Exclusive of cold-junction errors.
Model DMM7510 condensed specifications
Model DMM7510 condensed specifications