10: Capturing and analyzing waveforms Model DMM7510 7½ Digit Multimeter
User's Manual
10-4 DMM7510-900-01 Rev. B / May 2015
Ripple noise on the output voltage
This test measures the ripple voltage captured on the output terminal of the buck converter evaluation
board.
For this test, you will:
Connect the 3 load resistor to the output terminals of the buck converter
Connect test leads to the output terminals of the buck converter
Supply 12 V to the input terminals of the buck converter
Reset the instrument
Select the digitize voltage function and set sample rate to 1 Msamples per second with a sample
count of 100
View the data the graph
Turn on horizontal cursors to measure voltage ripple magnitude
Device connections
The connections for the ripple noise test are shown in the figure below.
Figure 35: Ripple noise on output voltage device connections