Instructions

Table Of Contents
9: Measuring power using digitizing and TSP-Link Model DMM6500 6½ Digit Multimeter
9-4 DMM6500-900-01 Rev. B / August 2019
Measuring power using digitizing and TSP-Link
For this application, you will:
Set the TSP-Link nodes for the DMM6500 instruments to 1 and 2.
Reset the instruments.
Initialize TSP-Link.
Configure the input and output TSP-Link triggers for the DMM6500 instruments.
Set up the digitize function.
Set up trigger models on both instruments.
Initiate measurements on both DMM6500 instruments.
Use buffer statistics to find the average current and voltage.
Find the average power usage of a Bluetooth device.
Print the results to the USER swipe screen.
Using SCPI commands
This example cannot be replicated in SCPI code because TSP-Link events and commands are only
available through the TSP command language.
However, you can use other instrument trigger interfaces, such as digital I/O or external trigger I/O, to
replace TSP-Link and achieve similar operation.
Setting up the nodes for TSP code
Before executing the TSP code, you must set up the nodes on the instruments and configure the
TSP-Link network.
To set up TSP-Link on the DMM6500:
1. Press the MENU key.
2. Under System, select Communication.
3. Select the TSP-Link tab.
4. Set the Node to 1 on the instrument that is used to measure voltage
5. Repeat these steps on the other instrument and set the Node to 2 on the instrument that is used
to measure current.
6. Select Initialize on each instrument.