Instructions
Table Of Contents
- Model DMM6500 6½-Digit Multimeter User's Manual
- 1 Introduction
- 2 Front-panel overview
- 3 Using a remote interface
- 4 Making basic front-panel measurements
- 5 Measuring DC voltage with high accuracy
- 6 Measuring 4-wire resistance with offset compensation
- 7 Scanning temperature at a set time interval
- 8 Grading and binning resistors
- 9 Measuring power using digitizing and TSP-Link
- 10 Troubleshooting FAQs
- About this section
- Where can I find updated drivers?
- Is there any software to help me get started?
- Why did my settings change?
- Why can't the DMM6500 read my USB flash drive?
- How do I upgrade the firmware?
- How do I change the command set?
- How do I save the present state of the instrument?
- How do I save what is displayed on the screen?
- What is the ethernet port number?
- 11 Next steps
- Contact information
Section
9: Measuring power using digitizing and TSP-Link Model DMM6500 6½ Digit Multimeter
User's Manual
9-2 DMM6500-900-01 Rev. B / August 2019
Equipment required
This application requires the following equipment:
• Two DMM6500 instruments
• Two KTTI-TSP communication and digital I/O accessory cards
• One computer set up for communication with the instrument
• One ethernet crossover cable
• Several insulated banana cables
• One device or component to be tested
Device connections
This application requires two KTTI-TSP communication and digital I/O accessory cards.
1. Insert one card into the accessory card slot on the rear panel of each instrument. Refer to
Install
the KTTI-TSP accessory (on page 3-16) card for installation instructions.
2. Connect the instruments with a crossover cable connected to the communication cards.
Figure 43: Two DMMs connected with TSP-Link
3. Connect the computer to the DMM6500 that is set to node one.
4. Connect the test leads of the instrument measuring voltage in parallel with the battery of the
device.
5. Connect the test leads of the instrument measuring current in series with the battery of the
device.