Instructions

Table Of Contents
7: Scanning temperature at a set time interval Model DMM6500 6½ Digit Multimeter User's
7-6 DMM6500-900-01 Rev. B / August 2019
Commands Descriptions
*RST
TRAC:CLE "defbuffer1"
TRAC:POIN 12960, "defbuffer1"
FUNC "TEMP", (@1:10)
TEMP:UNIT CELS, (@1:10)
TEMP:TRAN CJC2001, (@1)
TEMP:TRAN TC, (@2:10)
TEMP:TC:TYPE K, (@2:10)
TEMP:TC:RJUN:RSEL EXT, (@2:10)
TEMP:NPLC 1, (@2:10)
ROUT:SCAN:INT 60
ROUT:SCAN:COUN:SCAN 1440
ROUT:SCAN:CRE (@1:10)
ROUT:SCAN:EXP "/usb1/scan24hr", SCAN, ALL
ROUT:SCAN:REST ON
INIT
*WAI
TRAC:DATA? 1, totalRdgs, "defbuffer1\", READ
Reset the DMM6500
Clear the data buffer
Set the scan count to 12,960
Set the function to temperature
Set up all channels to use degrees
Celsius
Set up channel 1 reference junction
Set the transducer type to thermocouple
Set thermocouple type to K
Set the reference junction to external
Set NPLC to 1
Set the scan-to-scan delay to 60 seconds
Set the scan count to 1440 (24 hr * 60 s)
Set the scan list
Prepare the scan to export the buffer to
the USB flash drive after each scan
Enable restart on power loss
Start the scan
Suspend the scan
Return data from the buffer after the scan
completes
Using TSP
The following TSP code is designed to be run from Keithley Instruments Test Script Builder (TSB).
TSB is a software tool that is available from tek.com/keithley
. You can install and use TSB to write
code and develop scripts for TSP-enabled instruments. Information about how to use TSB is in the
online help for TSB and in the “Introduction to TSP operation” section of the Model DMM6500
Reference Manual.
To use other programming environments, you may need to make changes to the example TSP code.
By default, the DMM6500 uses the SCPI command set. You must select the TSP command set
before sending TSP commands to the instrument.
To enable TSP commands:
1. Press the MENU key.
2. Under System, select Settings.
3. Set the Command Set to TSP.
4. At the prompt to reboot, select Yes.