Datasheet

TEK.COM 17
DMM6500 6½-Digit Bench/System Digital Multimeter
System Specifications
Typical Reading Rates, DC Functions
37, 3 8
60 Hz (50 Hz) Operation
NPLC
Functions: DCV (10 V)
2-wire (10 k), DCI (1 mA)
Functions: 4-wire (1 k)
4-wire and 3-wire RTD Function: Thermistor or Thermocouple
Measurements (readings per second)
39
Buffer Computer Buffer Computer Buffer Computer
5 12 (10) 11 ( 9) 5 (4) 5 (4) 12 (10) 11 (9)
1 59 (48) 58 (48) 28 (23) 28 (23) 59 (49) 57 (48)
0.1 584 (490) 440 (380) 180 (160) 170 (150) 580 (480) 440 (380)
0.01 4900 (4100) 4800 (4100) 400 (390) 400 (390) 4800 (4100) 4700 (4000)
0.0005 20600 (20600) 19800 (19800) 460 (460) 460 (460) 21000 (21000) 20300 (20300)
Typical Reading Rates, AC Functions
37
60 Hz (50 Hz) Operation
Function: ACV, ACI Function: Frequency, Period
Measurements
(readings per second)
Detector Bandwidth Aperture Buffer or Computer
3 Hz 200 ms 1
30 Hz 20 ms 10
300 Hz 2 ms 100
Scanning/Multiple Channels (with optional scan cards)
40
Typical Scanning Measurement Rates Measurements Into Buffer/Computer (channel per second)
Scanning DCV or 2-wire >90 with 2000-SCAN card, >90 with 2001-TCSCAN card
Scanning Thermocouple, Thermistor, or 2-wire RTD >85 with 2000-SCAN card, >85 with 2001-TCSCAN card
Scanning 4-wire and 3- or 4-wire RTD >80 with 2000-SCAN card, >80 with 2001-TCSCAN card
Scanning ACV >60 with 2000-SCAN card, >60 with 2001-TCSCAN card
Scanning Alternating DCV and 2-wire >85 with 2000-SCAN card, >85 with 2001-TCSCAN card
Notes
37. Reading speeds for autozero off, fixed range, autodelay off, offset compensation off, and open lead detector off where applicable.
38. Buffer measurements: For <0.1 PLC, multisample, and single buffer transfer binary reading only.
39. Computer measurements: For 5 PLC, 1 PLC, and 0.1 PLC single reading and single transfer to computer (USB).
40. Set-up conditions of the factory default setting with the following exceptions: 3.5 digits (0.0005 PLC), autorange off, autozero off, autodelay off, and open lead detection off.