Datasheet
TEK.COM 15
DMM6500 6½-Digit Bench/System Digital Multimeter
Diode
Diode Voltage Accuracy ±(% of reading + additional uncertainty)
32
Voltage Measure
Range Resolution
Maximum Voltage
Measurement Test Current (±5%)
2 Year Accuracy
T
CAL
±5°C
Temperature
Coefficient
10 V 10 μV
12 V 10 μA 0.0045 + 60.0 μV 0.0008 + 10 μV
10 V 100 μA 0.0045 + 80.0 μV 0.0008 + 10 μV
7 V 1 mA 0.0045 + 170.0 μV 0.0010 + 10 μV
7 V 10 mA 0.0045 + 1.1 mV 0.0010 + 10 μV
Digitize
Digitize DC Voltage Accuracy ±(% of reading + % of range)
33
Range Resolution Input Impedance
2 Year Accuracy
T
CAL
±5°C Temperature Coefficient
100 mV 10 μV >10 GΩ or 10 MΩ ±1% 0.040 + 0.020 0.0025 + 0.0030
1 V 100 μV >10 G Ω or 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010
10 V 1 mV >10 G Ω or 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010
100 V 10 mV 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010
1000 V 100 mV 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010
Digitize DC Current Accuracy ±(% of reading + % of range)
33
Range Resolution Burden Voltage
2 Year Accuracy
T
CAL
±5°C Temperature Coefficient
100 μA 10 nA <0.14 V 0.07 + 0.05 0.0030 + 0.0035
1 mA 100 nA <0.17 V 0.07 + 0.03 0.0030 + 0.0035
10 mA 1 μA <0.17 V 0.05 + 0.03 0.0030 + 0.0035
100 mA 10 μA <0.20 V
34
0.05 + 0.03 0.0020 + 0.0035
1 A 100 μA <0.55 V
34
0.07 + 0.03 0.0040 + 0.0035
3 A 100 μA <1.70 V
34
0.09 + 0.04 0.0040 + 0.0035
10 A 1 mA <0.50 V 0.25 + 0.08 0.0060 + 0.0100
Notes
32. Specifications do not include errors that may arise from user’s cable or connection resistance.
33. DC accuracy specified with 1000 samples per second, 100-reading digital filter.
34. When using the rear terminals, add 0.1 V to the 100 mA range and 0.5 V to the 1 A and 3 A ranges.