Datasheet

Datasheet
TEK.COM34
The 7707 plug-in module offers 10 channels of 2-pole or 5 channels
of 4-pole multiplexer switching that can be configured as two
independent banks of multiplexers. The 7707 also provides 32 digital
input/output channels (four 8-bit ports) for I/O control. Connect the
7707 to industry standard solid-state relays to switch up to 980 VA.
Key Features
300 V, 1 A capacity; 60 W, 125 VA maximum (analog)
33 V, 100 mA capacity (digital)
Digital outputs are short circuit protected
Relay closures stored in onboard memory
Ordering Information
7707 32-channel Digital I/O Module with 10-channel
Differential Multiplexer
Series 7700 Plug-In Switch Modules
Channel 12 Channel 14
HI
LO
Sense
HI
LO
Input
To
Internal
DMM
Channel 17
(see Note)
Backplane
isolation
Channel 16
(see Note)
Backplane
isolation
NOTES:
Channels 15—17 in this schematic refer
to the designations used for control
and not actual available channels.
For more information, refer to the
ROUTe:MULT command section in the
Instrument Users Manual.
Bit
1—16
Channel 11
DIO
0
1
2
3
4
5
6
7
0
1
2
3
4
5
6
7
Digital I/O
Bit
17—32
Channel 13
DIO
0
1
2
3
4
5
6
7
0
1
2
3
4
5
6
7
Digital I/O
Channel 1
HI
LO
Channel 5
HI
LO
(Channels 2—4)
Channel 6
HI
LO
Channel 10
HI
LO
(Channels 7—9)
Channel 15
2-Pole (Open)
4-Pole (Closed)
(see Note)
Card Sense
HI
LO
Card Input
HI
LO
7707 32-channel Digital I/O Module
with 10-channel Differential Multiplexer
Datasheet
Specifications
Capabilities
Channels 1–10 Multiplex one of 10 2-pole or one of 5 4-pole signals into DMM.
Channels 11–14 32 Digital Inputs/Outputs referenced to chassis ground.
Thermal Protection Channels 11–14 are thermally protected to 1 A up to 25 V.
Inputs (Channels 1–10)
Maximum Signal Level: Any Channel to Any Channel (1–10): 300 VDC or 300 Vrms (425
V peak) for AC waveforms, 1 A switched, 60 W, 125 VA maximum.
Contact Life (typ.) >10
5
operations at max. signal level
>10
8
operations no load
1
.
1. Minimum signal level 10 mV, 10 μA.
Contact Resistance <1 Ω any path and additional 1 Ω at end of contact life.
Contact Potential <6 μV typical per contact pair and additional 5 μV with Channels
11–14 at rate V
OUT
(L).
Offset Current <100 pA.
Connector Type: 50-pin male D-shell, Channels 11–14. 25-pin female D-shell,
Channels 110. Supplied with female and male IDC ribbon cable
connectors.
Isolation Between Any Two Terminals
>10
9
Ω, <100 pF with isolation channels 16 and 17 open.
Isolation Between Any Terminal and Earth: >10
9
Ω, <200 pF.
Cross Talk (1 MHz, 50 Ω Load): <–35 dB.
Insertion Loss (50 Ω Source, 50 Ω Load)
<0.1 dB below 1 MHz. <3 dB below 2 MHz.
Common Mode Voltage: 300 VDC or 300 Vrms (425 V peak) for AC
waveforms between any terminal and chassis.
Digital Input/Output (Channels 11–14)
V
IN
(L): <0.8 V (TTL). V
IN
(H): >2 V (TTL).
V
out
(L): <1.0 V @ I
out
= 100 mA. V
out
(H): >2.4 V @ I
out
= 1 mA.
V
out
(H)MAX.: <40 V with external open drain pull-up.
Read/Write Speed 50/s.
General
10 Channels 10 channels of 2-pole relay input. All channels configurable to
4-pole.
Relay Type Latching electromechanical.
Actuation Time <3 ms.
Capacity 2700: (1) 7707 and (1) 77XX, except 7706.
2701: Any combination of 77XX modules.
2750: (4) 7707 and (1) 77XX, except 7706. A 7706 module may be
substituted for a 7707 module.
Environmental Operating Environment: Specified for 0° to 50°C. Specified to
80% R.H. at 35°C.
Storage Environment:25° to 6C.
EMC: Conforms to European Union EMC Directive.
Safety: Conforms to European Union Low Voltage Directive
RoHS: Conforms to European Union RoHS Directive
Warranty: 1 year
Weight <0.5 kg (1.16 lb).
Supplied Accessories
7707-306A 50-pin D-Sub Female IDC Connector Kit
7709 -307A 25-pin D-Sub Male IDC Connector Kits
Available Accessories
7790 50-pin Male, 50-pin Female, 25-pin Male D-shell IDC Connectors
7705-MTC-2 50 Pin Male to Female D-sub Cable, 2 m (6.6 ft).
7707-MTC-2 25 Pin Male to Female D-sub Cable, 2 m (6.6 ft)..
Available Services
7707-3Y-E W 1-year factory warranty extended to 3 years from date of shipment