Datasheet
Datasheet
TEK.COM24
Remote Interface: Standard
LAN/LXI Compliance RJ-45 Connector: 10/100BT. IP Configuration: Static or DHCP (manual or automatic).
Web Interface: Virtual front panel. LXI Compliance: LXI version 1.4 core 2016.
USB Device (rear panel, Type B) 2.0 full speed, USBTMC compliant
USB Host (front panel, Type A) USB 2.0, support for flash drives, FAT32. Capability: Import/export instrument configuration files, reading
buffers, screen captures, and scripts
Language
SCPI (default) Default command set, Standard Commands for Programmable Instruments, SCPI-1999
TSP Embedded Test Script Processor (TSP) accessible from any host interface; responds to high-speed test
scripts comprised of remote commands and statements (for example, branching, looping, and math); able
to execute test scripts stored in memory without host intervention
Emulation Modes Keithley Model 2700 and Model 2701
Math Functions
REL, Minimum, Maximum, Average, Standard Deviation, peak-peak, dB, Limit Test, Percent, 1/x, and
mX+b with user-defined units displayed
Miscellaneous
Real-time Clock Lithium battery backup, CR2032 coin-type, factory replaceable, (3+ years of battery life); set and read year,
month, day, hour, minute, and second. (Note: Seconds are not adjustable.)
Timestamp Resolution 15 ns with standard or full buffer style
Password Protection 30 characters
Alarms Up to six: see Optional Interfaces and Programmable Digital I/O
Power Failure Recovery Mode User selectable, resumes scanning once power is re-applied
Optional Interfaces and Programmable Digital I/O
KTTI-GPIB GPIB IEEE-488.1 compliant; supports IEEE-488.2 common commands and status model topology
KTTI-RS232 RS232, 9-pin d-sub female connector; standard baud rates from 300 to 115,200 bps are supported
KTTI-TSP RJ-45 (quantity 2); TSP-Link
®
expansion interface allows TSP-enabled instruments to trigger and
communicate with each other
Digital I/O For KTTI-RS232, KTTI-GPIB, and KTTI-TSP
Connector: 9 pin d-sub female
5 V Power Supply Pin: Limited to 500 mA > 4 V (solid-state fuse protected)
Lines: Six input / output, user-defined for control, alarms (limits), or triggering
Input Signal Levels: 0.7 V (maximum logic low), 3.7 V (minimum logic high)
Input Voltage Limits: –0.25 V (absolute minimum), 5.25 V (absolute maximum)
Maximum Source Current: 2.0 mA at > 2.7 V (per pin)
Maximum Sink Current: –50 mA at 0.7 V (per pin, solid state fused)