Datasheet
Datasheet
TEK.COM22
Typical Function and Range Change Speed
Function
Function Change
Time
46
Range Change
Time
47
Autorange
Time
46
DCV, DCI, or 2-wire Ω
48
<4 ms <1.3 ms
<3.2 ms
4-wire Ω
49
or 3-wire RTD <5.5 ms
Thermistor —
Frequency or Period (2 ms aperture)
<1800 ms <50 ms
50
<50 ms
50
ACV (300 Hz bandwidth)
ACI (300 Hz bandwidth) <100 ms <4 ms <5 ms
Capacitance <4 ms <3 ms <30 ms
Digitize <4 ms <5 ms —
Diode <11 m s — —
Continuity <11 m s — —
Thermocouple <4 ms — —
Bus Transfer Speed
51
USB LAN GPIB RS232 (Baud 115200)
Average for 1000 readings
(binary)
441,000 268,000 201,000 10,000
Average for 1000 readings with
relative timestamp (binary)
272,000 150,000 105,000 2,900
Average for 1000 readings with
formatted elements
52
46,000 29,000 17,000 290
Typical Digitize Voltage or Current
53
Sampling Rate
Measurements over USB to Computer
(readings per second)
10 kS/s Up to 10,000
50 kS/s Up to 50,000
100 kS/s Up to 100,000
1 MS/s up to 7 s maximum duration At least 90,000
Triggering
Trigger Sources Front panel trigger key, timer, command interface, LAN/LXI, Trigger In (BNC rear panel), Digital I/O (optional
accessory card), and TSP-Link
®
(optional accessory card)
External Trigger Delay <1 µs when triggering from accessory card or rear BNC input
External Trigger Jitter <1 µs when triggering from accessory card or rear BNC input
External Trigger In/Trigger Out 0 V to 5 V logic signal input and output, TTL-compatible, programmable edge pulse
Minimum pulse width: 1 µs
External Trigger Out, Maximum Rate
Up to 90 kHz, measurement dependent
External Trigger In, Maximum Rate Up to 150 kHz, measurement dependent
Notes
46. 3.5 digits, autozero off, 0.0005 PLC, excludes measurement time.
47. DCV = 10 V; 2-wire or 4-wire = 1 kΩ; DCI = 1 mA; ACI = 1 mA; ACV = 1 V; Capacitance = 10 µF.
48. 2-wire function for 100 Ω range and up. For the 10 Ω range, add 2.7 ms.
49. 4-wire function for 100 Ω range and up. For the 1 Ω and 10 Ω ranges, add 2.7 ms.
50. When ranging to 10 V and above, add 1.8 s.
51. SCPI programmed using 4-byte binary format.
52. Format elements: Reading, relative timestamp, channel, and unit.
53. SCPI programmed using 4-byte binary format.