Datasheet

TEK.COM 19
DAQ6510 Data Acquisition and Logging, Multimeter System
Diode
Diode Voltage Accuracy ±(% of reading + additional uncertainty)
35
Voltage Measure
Range Resolution
Maximum Voltage
Measurement Test Current (±5%)
2 Year Accuracy
T
CAL
±5°C
Temperature
Coefficient
10 V 10 µV
12 V 10 µA 0.0045 + 60.0 µV 0.0008 + 10 µV
10 V 100 µA 0.0045 + 80.0 µV 0.0008 + 10 µV
7 V 1 mA 0.0045 + 170.0 µV 0.0010 + 10 µV
7 V 10 mA 0.0045 + 1.1 mV 0.0010 + 10 µV
Digitize
Digitize DC Voltage Accuracy ±(% of reading + % of range)
36
Range Resolution Input Impedance
2 Year Accuracy
T
CAL
±5°C Temperature Coefficient
100 mV 10 µV >10 GΩ or 10 MΩ ±1% 0.040 + 0.020 0.0025 + 0.0030
1 V 100 µV >10 GΩ or 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010
10 V 1 mV >10 G Ω or 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010
100 V 10 mV 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010
1000 V 100 mV 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010
Digitize DC Current Accuracy ±(% of reading + % of range)
36
Range Resolution Burden Voltage
2 Year Accuracy
T
CAL
±5°C Temperature Coefficient
100 µA 10 nA <0.14 V 0.07 + 0.05 0.0030 + 0.0035
1 mA 100 nA <0.17 V 0.07 + 0.03 0.0030 + 0.0035
10 mA 1 µA <0.17 V 0.05 + 0.03 0.0030 + 0.0035
100 mA 10 µA <0.20 V 0.05 + 0.03 0.0020 + 0.0035
1 A 100 µA <0.55 V
37
0.07 + 0.03 0.0040 + 0.0035
3 A 100 µA <1.70 V
37
0.09 + 0.04 0.0040 + 0.0035
Notes
35. Specifications apply to the front-panel inputs, only. Specifications do not include errors that may arise from user’s cable or connection resistance.
36. DC accuracy specified with 1000 samples per second, 100-reading digital filter.
37. Add 1.5 V for 1 A and 3 A ranges when used with a plug-in module.