Datasheet

Datasheet
TEK.COM16
Temperature Characteristics
Thermocouple Conversion ITS-90
Thermocouple Reference Junction External (CJC on 2001-TCSCAN or user-provided with 2000-SCAN) or simulated (fixed)
Open Thermocouple Detection Selectable per channel (open >130 kΩ; default on.
Earth Isolation 500 V
PEAK
> 0 GΩ and <300 pF any terminal to chassis
AC Voltage
AC Voltage Accuracy ±(% of reading + % of range)
19
Range Resolution
Calibration
Cycle 3 Hz to 5 Hz 5 Hz to 10 Hz
10 Hz to
20 kHz
20 kHz to
50 kHz
50 kHz to
100 kHz
100 kHz to
300 kHz
100 mV 100 nV
1 V 1 µV
10 V 10 µV
100 V 100 µV
750 V 100 µV
24 hours 1.00 + 0.02 0.35 + 0.02 0.04 + 0.02 0.10 + 0.04 0.55 + 0.08 4.00 + 0.50
90 days 1.00 + 0.03 0.35 + 0.03 0.05 + 0.03 0.11 + 0.0 5 0.60 + 0.08 4.00 + 0.50
1 year 1.00 + 0.03 0.35 + 0.03 0.06 + 0.03 0.12 + 0.05 0.60 + 0.08 4.00 + 0.50
2 years 1.00 + 0.03 0.35 + 0.03 0.07 + 0.03 0.13 + 0.05 0.60 + 0.08 4.00 + 0.50
Temperature Coefficient
0.100 + 0.003 0.035 + 0.003 0.005 + 0.003 0.011 + 0.005 0.060 + 0.08 0.200 + 0.020
AC Voltage Characteristics
Overrange (voltages in V
RMS
) 20% on 100 mV, 1 V, 10 V, and 100 V ranges. 0% for 750 V range.
AC Measurement Method AC-coupled digital sampling with anti-alias filter
Crest Factor (excludes sine wave) Crest factors of up to 3:1 at full-scale input or 10:1 maximum, whichever is greater
Autorange selects optimum range for crest factor up to 10:1
Accuracy specifications apply to all crest factors and are limited to a product of
(crest factor) × (fundamental frequency) 3 kHz
Volt*Hertz Product 8 × 10
7
V*Hz
21
Common Mode Rejection Ratio >70 dB, for 1 kΩ unbalance in LO lead
Detector Bandwidth Setting of 3 Hz, 30 Hz, or 300 Hz sets maximum measurement aperture of 200 ms, 20 ms, or 2 ms,
respectively; only signals with frequency greater than the detector bandwidth are measured.
Input Impedance 1.1 M Ω ±2%, in parallel with <100 pF
Input Protection Front-panel input: 1100 VPEAK or 400 VDC on all ranges
Rear-panel input through the plug-in modules: 600 V
PEAK
on all ranges (note this level exceeds the
maximum input signal for modules; see Maximum Input Signal Levels for 7700 Series plug-in modules
listed below.
Maximum DCV 400 V on any ACV range
ACV Frequency Frequency reading automatically returned in reading buffer when in full buffer mode.
Frequency readings are specified as in the frequency and period table.
Plug-in Module Maximum Input Signal Levels
Module Maximum Input Signal Level
7700, 7702, 7703, 7706, 7707, 7708, 7709 300 V
RMS
/ 425 V
PEAK
7701 150 V
RMS
/ 212 V
PEAK
7710 42 V
RMS
/ 60 V
PEAK
Notes
20. Specifications are for sine wave inputs >5% of range.
21. Guaranteed by design.