Datasheet
TEK.COM 15
DAQ6510 Data Acquisition and Logging, Multimeter System
Temperature
Thermocouple Accuracy ±°C
17
Type Resolution Range
2 Year Accuracy
T
CAL
±5°C; all uncertainties in °C
Temperature
Coefficient
in°C/°C
Simulated or External CJC Internal CJC (on module)
Front Terminals,
7700, 7702, 7708
7706,
7709,
7710
7701,
7703,
7707
7700,
7708 7706 7710
J 0.001°C
0 °C to 760 °C 0.20 0.20 0.20 1.0 0 1.0 0 1.30 0.03
–200 °C to < 0 °C 0.20 0.20 0.40 1.50 1.9 0 3.00 0.03
K 0.001°C
0 °C to 1372 °C 0.20 0.20 0.30 1.0 0 1.0 0 1.00 0.03
–200 °C to < 0 °C 0.30 0.30 0.50 1.70 2.00 2.00 0.03
N 0.001°C
0 °C to 1300 °C 0.20 0.20 0.30 1.00 1.00 1.40 0.03
–200 °C to < 0 °C 0.50 0.60 0.80 1.80 2.30 3.60 0.03
T 0.001°C
0 °C to 400 °C 0.20 0.20 0.20 1.00 1.00 1.50 0.03
–200 °C to < 0 °C 0.30 0.30 0.50 1.60 2.00 3.50 0.03
E 0.001°C
0 °C to 1000 °C 0.20 0.20 0.20 1.00 1.0 0 1.20 0.03
–200 °C to < 0 °C 0.20 0.30 0.30 1.50 1.80 3.00 0.03
R 0.010 °C
600 °C to 1768 °C 0.40 0.50 0.70 1.00 1.00 1.20 0.03
0 °C to < 600 °C 0.80 1.00 1.50 1.50 1.60 2.20 0.03
S 0.010 °C
600 °C to 1768 °C 0.40 0.50 0.70 1.00 1.00 1.20 0.03
0 °C to < 600 °C 0.80 1.00 1.50 1.30 1.60 2.20 0.03
B 0.010 °C
1100 °C to 1820 °C 0.40 0.50 0.80 1.00 1.00 1.0 0 0.03
350 °C to < 1100 °C 1.20 1.50 2.20 1.10 1.40 1.50 0.03
Resistance Temperature Detector (RTD) Accuracy ±°C
Types: 100 Ω platinum PT100, D100, F100, PT385, and PT3916 or user-configurable 0 Ω to 10 kΩ
Measurement
Method Resolution Range
2 Year Accuracy
T
CAL
±5°C
Temperature Coefficient
in°C/°C
2-wire
18
0.01°C –200° to 850°C 0.80 0.003
3-wire
19
0.01°C
–200° to 600°C 0.35 0.003
>600° to 850°C 0.37 0.003
4–wire 0.01°C
–200° to 600°C 0.06 0.003
>600° to 850°C 0.12 0.003
Thermistor Accuracy ±°C
Types: 2.2 kΩ, 5 kΩ, and 10 kΩ
Measurement
Method Resolution Range
2 Year Accuracy
T
CAL
±5°C
Temperature Coefficient
in°C/°C
2-wire 0.01°C –80° to 150°C 0.08 0.002
For readings >70°C, add this additional uncertainty per Ω of lead, channel, and contact resistance
Thermistor Type Common Model Number 70° to 100°C >100° to 150°C
2.2 kΩ 44004 0.22°C per Ω 1.11° C p e r Ω
5 kΩ 44007 0.10°C per Ω 0.46°C per Ω
10 kΩ 44006 0.04°C per Ω 0.19°C per Ω
Notes
17. Accuracy excludes probe errors.
18. Specifications do not include errors that may arise from user’s cable or terminal resistance.
19. 3-wire RTD accuracy is for <0.1 Ω lead-resistance mismatch for input HI and LO. Add 0.25°C per 0.1 Ω of HI-LO resistance mismatch.