Datasheet
www.keithley.com
1.888.KEITHLEY (U.S. only)
A GREATER MEASURE OF CONFIDENCE
Integra Series Specifications
SWITCH/MEASURE SYSTEMS
21
Integra Series Multimeter/Switch Systems
Mainframes
DC NOTES
1. 20% overrange except on 1000V and 3A.
2. Add the following to “ppm of range” uncertainty; 100mV 15ppm; 1V and 100V 2ppm; for Model 2750 1Ω and Dry Circuit Ω
40ppm; 10→1MΩ 2ppm, for Models 2700/2701 100Ω 30ppm, 20mA and 1A 10ppm, 100mA 40ppm.
3. ±2% (measured with 10MΩ input resistance DMM, >10GΩ DMM on 10MΩ and 100MΩ ranges). For Dry Circuit Ω, ±25% with
Input HI connected to Sense HI; with Sense HI disconnected add 30mV.
4. Relative to calibration accuracy.
5. For signal levels >500V, add 0.02ppm/V uncertainty for portion exceeding 500V.
6. Specifications are for 4-wire Ω, 1Ω, 10Ω, and 100Ω with offset compensation on. With 77XX plug-in modules, LSYNC on. With off-
set compensation on, OPEN CKT. VOLTAGE is 12.8V. For 2-wire Ω add 1.5Ω to “ppm of range” uncertainty. 1Ω range is 4-wire only.
7. Must have 10% matching of lead resistance in Input HI and LO.
8. Add the following to “ppm of reading” uncertainty when using plug in modules:
9. Add 1.5V when used with plug in modules.
10. For RATIO, DCV only. For AVERAGE, DCV and Thermocouples only. Available with plug in modules only.
11. Add 6µV to “of range” uncertainty when using Models 7701, 7703, and 7707, and 3µV for Models 7706 and 7709.
12. Auto zero off.
13. For LSYNC On, line frequency ±0.1 %. For LSYNC Off, use 60dB for ≥ 1PLC.
14. For 1kΩ unbalance in LO lead. AC CMRR is 70dB.
15. Speeds are for 60Hz (50Hz) operation using factory defaults operating conditions (*RST). Autorange off, Display off, Limits off,
Trigger delay = 0.
16. Speeds include measurements and binary data transfer out the GPIB or ASCII data transfer for Ethernet and RS-232 (reading ele-
ment only).
17. Sample count = 1000, auto zero off (into memory buffer).
18. Auto zero off, NPLC = 0.01 (Models 2700 and 2750), NPLC = 0.002 (Model 2701).
DC SPEED vs. NOISE REJECTION
RMS Noise
10V Range
Rate Filter Readings/s
12
Digits 2700,2750 2701 NMRR CMRR
14
10 50 0.1 (0.08) 6.5 < 1.2 µV < 2.5 µV 110 dB
13
140 dB
1 Off 15 (12) 6.5 < 4 µV < 6 µV 90 dB
13
140 dB
0.1 Off 500 (400) 5.5 < 22 µV < 40 µV — 80 dB
0.01 Off 2500 (2000) 4.5 < 150 µV < 300 µV — 80 dB
0.002 Off 3500 (2800) 3.5 — < 1 mV — 60 dB
DC MEASUREMENT CHARACTERISTICS
DC VOLTS
A-D LINEARITY: 2.0 ppm of reading + 1.0 ppm of range.
INPUT IMPEDANCE:
100mV–10V Ranges: Selectable >10GΩ // with <400pF or 10MΩ ±1%.
100V, 1000V Ranges: 10MΩ ±1%.
Dry Circuit: 100kΩ ±1% // <1µF.
EARTH ISOLATION: 500V peak, >10GΩ and <300pF any terminal to chassis.
INPUT BIAS CURRENT: <75pA at 23°C.
COMMON MODE CURRENT: <500nApp at 50Hz or 60Hz.
AUTOZERO ERROR: Add ±(2ppm of range error + 5µV) for < 10 minutes and ±1°C.
INPUT PROTECTION: 1000V, all ranges. 300V with plug in modules.
RESISTANCE
MAXIMUM 4WΩ LEAD RESISTANCE: 80% of range per lead (Dry Ckt mode). 5Ω per
lead for 1Ω range; 10% of range per lead for 10Ω, 100Ω, and 1kΩ ranges; 1kΩ
per lead for all other ranges.
OFFSET COMPENSATION: Selectable on 4WΩ, 1Ω, 10Ω, 100Ω, 1kΩ, and 10kΩ
ranges.
CONTINUITY THRESHOLD: Adjustable 1 to 1000 Ω
INPUT PROTECTION: 1000V, all Source Inputs, 350V Sense Inputs. 300V with plug-in
modules.
DC CURRENT
SHUNT RESISTORS: 100mA–3A, 0.1Ω. 20mA, 5Ω.
INPUT PROTECTION: 3A, 250V fuse.
THERMOCOUPLES
CONVERSION: ITS-90.
REFERENCE JUNCTION: Internal, External, or Simulated (Fixed).
OPEN CIRCUIT CHECK: Selectable per channel. Open >11.4kΩ ±200Ω.
10 kΩ 100 kΩ 1 MΩ 10 MΩ 100 MΩ
All Modules:
220 ppm 2200 ppm
7701, 7703, 7707, 7709 Modules:
10 ppm 100 ppm 1000 ppm 1% 10%
7706, 7708, 7710 Modules:
5 ppm 50 ppm 500 ppm 5000 ppm 5%
7710 Module 23°C ±5°C:
11 ppm 110 ppm 1100 ppm 1.1% 11%
19. Additional Uncertainty:
Plug-In Modules
Type Range
Front Terminals
Sim. Ref.
Junction
7709 Sim.
Ref. Junction
7701/03/07
Sim. Ref.
Junction
7700/08
Using
CJC
7706
Using
CJC
7710
Using
CJC
J –200° to 0°C 0.1 0.1 0.3 0.8 1.6 4.5
K –200° to 0°C 0.2 0.2 0.4 0.8 1.6 1.0
N –200° to 0°C 0.3 0.3 0.6 0.8 1.6 2.5
T –200° to 0°C 0.2 0.1 0.4 0.8 1.6 2.5
E –200° to 0°C – 0.1 0.3 0.8 1.6 2.5
R 0° to +400°C 0.4 0.6 1.2 0.5 1.0 2.2
S 0° to +400°C 0.4 0.6 1.2 0.5 1.0 2.2
B +350° to +1100°C 0.8 0.3 1.7 0.5 1.0 2.2
Type Range 7710 Using CJC
J 0° to +760°C 1.5
K 0° to +1372°C –
N 0° to +1300°C 0.5
T 0° to +400°C 0.5
E 0° to +1000°C 0.5
R +400° to +1768°C 0.9
S +400° to +1768°C 0.9
B +1100° to +1820°C 0.9
20. For lead resistance >0Ω, add the following uncertainty/Ω for measurement temperatures of:
21. For 4-wire Ω only, offset compensation on, LSYNC on.
22. For Dry Circuit 1kΩ range, 2 readings/s max.
23. For 2750 Front Inputs, add the following to Temperature Coefficient “ppm of reading” uncertainty: 1MΩ 25ppm, 10MΩ 250ppm,
100MΩ 2500ppm. Operating environment specified for 0°C to 50°C and 50% RH at 35°C.
24. Model 2750 only.
25. Front panel resolution is limited to 0.1Ω.
70°–100°C 100°–150°C
2.2 kΩ
(44004) 0.22°C 1.11°C
5.0 kΩ
(44007) 0.10°C 0.46°C
10 kΩ
(44006) 0.04°C 0.19°C