5-16 Delta, Pulse Delta, and Differential Conductance Model 6220/6221 User’s Manual
Return to Section 5 topics
Operation
Delta operation – front panel
The system configuration for front panel stand-alone operation is shown in 
Figure 5-2A on page 5-5.
1. Connections are shown in the following illustrations. All power must be 
removed from all components in the system before making connections: 
• Figure 5-3, page 5-6 – System connections
• Figure 5-5, page 5-8 – DUT test connections
2. Configure communications for the Models 2182/2182A and 622x as 
explained in “Configuring communications,” page 5-9.
3. On the Model 2182/2182A, select the desired measurement range (using 
the RANGE keys) and the integration rate (using the RATE key). Rate must 
be set to an integer value (1, 2, 3, ... up to 50 or 60). If some other rate is 
selected, it will automatically be changed to 1PLC by the Model 622x during 
the arming process.
4. On the Model 622x, press CONFIG and then DELTA to access the 
CONFIGURE DELTA menu. For details on these settings, see “Configura-
tion settings” on page 5-15.
a. Set I-HIGH, I-LOW, DELAY, COUNT, and COMPLIANCE ABORT.
b. When finished, use the EXIT key to back out of the menu structure.
5. Set the measurement units.
The basic Delta reading is in volts (which is the default setting). However, it 
can instead be expressed (and displayed) as an Ohms, Watts, or Siemens 
reading. See “Delta” on page 5-13 for details. 
To set the measurement units, press the UNITS key to display the 
READING UNITS menu. Select VOLTS, OHMS, WATTS, or SIEMENS. 
Measurement units can be changed while Delta is running.
6. On the Model 622x, press the DELTA key to arm Delta. 
The Model 6221 is armed when the message “DELTA ARMED Press TRIG 
to start” is displayed briefly and the ARM annunciator turns on. 
7. On the Model 622x, press the TRIG key to start taking Delta readings and 
send them to the buffer. 
If a finite Delta count is being used, the Delta measurements will stop after 
the last Delta measurement is performed. However, Delta remains armed 
and can be run again by pressing the TRIG key. The new Delta readings 
will overwrite the old readings in the buffer.
If the infinite Delta count is being used, Delta will run continuously. If the 
buffer fills, Delta readings will stop being stored, even though Delta 
continues to run. 