Model 6220/6221 User’s Manual Delta, Pulse Delta, and Differential Conductance 5-41
Return to Section 5 topics
A. Trace points specifies the size of the buffer. Buffer size should be the same 
value as the number of Differential Conductance readings in the test. See 
Section 6 for details on all buffer commands.
B. The initiate command starts Differential Conductance readings. After all 
Differential Conductance readings are performed, Differential Conductance 
will stop running. At this point another initiate command will re-start Differ-
ential Conductance. New Differential Conductance readings will overwrite 
the old Differential Conductance readings in the Model 622x buffer. 
6. Read Differential Conductance readings – While Differential Conduc-
tance is running, the latest Model 2182/2182A reading can be read by the 
Model 622x using the following command:
The above read command reads the last Differential Conductance reading 
that was performed by the Model 2182/2182A. If this command is sent 
before a new reading is available, the last Differential Conductance reading 
will again be returned. 
7. When finished with Differential Conductance, it can be disarmed by send-
ing the following command:
8. Recall stored Differential Conductance readings – Model 2182/2182A 
Differential Conductance readings were sent to the buffer of the Model 
622x. Send the following read command to read the buffer:
Setup and arm commands
Commands sent from the PC to the Model 622x to set up and arm Differential 
Conductance are listed in Table 5-4. 
SENS:DATA?
‘ Reads the latest reading.
NOTE If a read command is sent when Differential Conductance 
is not running, error -221 Settings Conflict will occur.
SOUR:SWE:ABOR ‘
‘
Stops Differential Conductance and places the 
Model 2182/2182A in the local mode.
TRACe:DATA? ‘ Read Differential Conductance buffer readings.