User Manual

Model 3760
10-Channel High Current Multiplexer Card User's Manual Section 1:
Introduction
3760-900-01 Rev. A / July 2017 1-3
Common mode voltage
500 V peak
Environmental conditions
Indoor use only
Operating: 0 °C to 50 °C (32 °F to 122 °F), 70% relative humidity up to
35 °C (95 °F); derate 3% relative humidity per °C, 35 °C to 50 °C,
noncondensing
Storage: −25 °C to 65 °C (-13 °F to 149 °F)
Altitude: 0 to 2000 m (0 to 6562 feet) above sea level
Pollution degree: 2
The 3760 is a high-current multiplexer card with the maximum capabilities of 500 V, 5 A or 100 VA
switching (resistive load only). Signal levels of this magnitude may be found in applications such as
power semiconductor testing. The 3760 can also be used in monitoring currents from DC power
supplies in general purpose automatic test equipment (ATE) bench or laboratory applications. For
easy input and output wiring the 3760 is designed with quick disconnect screw terminals. Each
channel has a HI and LO connection with a common guard. Each of the ten channels consist of a
single 2 Form C relay.
Figure 1: Model 3760 simplified schematic
The card is field installable in an appropriate Keithley Series 3700A Systems Switch/Multimeter (for
instance, the 3706A) using the mainframe’s rear panel. Channel switching is accomplished in less
than 15 mS (exclusive of mainframe) and the specified relay life is greater than 10
7
closures (greater
than 10
5
closures at maximum signal levels). With the 3760’s high channel isolation (>10
9
Ω), it is well
insulated from cross channel noise and interference. The 3760’s high input isolation (>10
7
Ω) reduces
loading to a minimum.
Specifications
For the most recent instrument specifications, search for the 3760 on the Keithley Instruments
website (http://www.tek.com/keithley).