Series 3700 System Switch/Multimeter Reference Manual  Section 15: Calibration 
3700S-901-01 Rev. C / July 2008  15-15 
Frequency calibration 
Make the connections as shown below, then perform the following calibration 
steps (AC Cal Step 17 and Step 18): 
Figure 15-7: Low frequency calibration 
AC Cal Step 17: 1V @ 10Hz (factory cal only) 
1.  Send the following commands: 
SEND dmm.func = dmm.ac_volts 
SEND dmm.range = 1 
2.  Source 1V @ 10Hz. 
3.  Send the following command: 
SEND dmm.calibration.ac(17,1) print("done") 
AC Cal Step 18: 1V @ 1kHz (factory cal only) 
Figure 15-8: Frequency verification 